5. Advanced User Guidebook
2. Regulate the probe attenuation.
In order to match the attenuation coefficient of the probe, it is required to adjust
the attenuation ration coefficient of the probe through the operating menu of the
Channel accordingly
(see “How to Set the Probe Attenuation Coefficient” on
P14). If the attenuation coefficient of the probe is 1:1, that of the oscilloscope
input channel should also be set to 1X to avoid any errors presented in the
displayed scale factor information and the measured data.
Take the Channel 1 as an example, the attenuation coefficient of the probe is 10:1,
the operation steps is shown as follows:
(1)
Press the
CH1 MENU
button, access CH1 SETUP menu.
(2)
Press the
H3
menu selection, the Probe menu will display at the right of the
screen, then press the
F2
button to select
10X
for the probe.
The Fig.5-3 illustrates the setting and the vertical scale factor when the probe of
the attenuation coefficient of 10:1.is used.
Fig. 5-3
Regulation of the Attenuation Ratio of the Probe
A List of the Attenuation Coefficient of Probes and the Corresponding Menu Settings:
Attenuation Coefficient of the Probe Corresponding Menu Setting
1:1 X1
10:1 X10
100:1 X100
1000:1 X1000
3. Setting of Waveform Inverted
Waveform inverted: the displayed signal is turned 180 degrees against the phase
of the earth potential.
User Manual of Smart Digital Storage Oscilloscope
23