Here are the procedures for carrying a self-calibration.
z
Disconnect all inputs, including probes and wires etc.
z
Press
UTILITY
to activate Function menu.
z
Press
F3
next to Do Self-Cal.
Now there is a yellow window popping up to ask for a confirmation. Press F3 again,
the Oscilloscope starts the self-calibration and yellow window remains the calibration
is done.
Please be ware that no contacts to any inputs channels till the calibration
is done.
SYS STAT (System State)
Press the
F1
menu selection button and choose “
SYS STAT
” item. The menu pops up
in the screen as Fig.47.
Fig.47
SYS STAT Menu
After entering into the SYS STAT menu, choose the corresponding function, with the
corresponding parameters shown in the screen. If press the
F1
menu selection button
and choose the function item
“Horizontal”,
the Horizontal System State will be
displayed in the screen. Press any other function button and exit from the SYS STAT
menu (see Fig.48).
Function Menu
Setting Description
Horizontal
Show the horizontal parameter of the channel.
Vertical
Show the vertical parameter of the channel.
Trigger
Show the parameters of the trigger system.
Misc
Show the serial number and edition number.