How to Implement the Probe Compensation
When connect the probe with any input channel for the first time, make this adjustment to
match the probe with the input channel. The probe which is not compensated or presents a
compensation deviation will result in the measuring error or mistake. For adjusting the
probe compensation, please carry out the following steps:
1.
Set the attenuation coefficient of the probe in the menu as 10X and that of the switch
in the probe as 10X
(see "
How to Set the Probe Attenuation Coefficient
connect the probe with the CH1 channel. If a probe hook tip is used, ensure that it
keeps in close touch with the probe. Connect the probe tip with the signal connector
of the probe compensator and connect the reference wire clamp with the ground wire
connector of the probe connector, and then push the
Autoset
button on the front
panel.
2.
Check the displayed waveforms and regulate the probe till a correct compensation is
achieved (see
and
Overcompensated
Compensated correctly
Under compensated
Figure 3-7 Displayed Waveforms of the Probe Compensation
3.
Repeat the steps mentioned if needed.
Figure 3-8 Adjust Probe
How to Set the Probe Attenuation Coefficient
The probe has several attenuation coefficients, which will influence the vertical scale
factor of the oscilloscope.
To change or check the probe attenuation coefficient in the menu of oscilloscope:
(1)
Push the function menu button of the used channels (
CH1
or
CH2 button
).
(2)
Select
Probe
in the bottom menu; select
Attenu
in the right menu, turn the
M
knob to
Содержание Dual-Channel Series
Страница 1: ...XDS3000 Dual Channel Series Digital Storage Oscilloscopes User Manual www owon com ...
Страница 29: ...3 Junior User Guidebook 22 Double Zoom Single Zoom ...
Страница 42: ...4 Advanced User Guidebook 35 Figure 4 2 Rectangle window Figure 4 3 Blackman window ...
Страница 43: ...4 Advanced User Guidebook 36 Figure 4 4 Hanning window Figure 4 5 Kaiser window ...