Owlstone Medical Breath Biospy ReCIVA-RD Breath Sampler Скачать руководство пользователя страница 4

 

©Owlstone Medical Ltd. 2018 

90-0796 Issue 2 OCT  2018 

Page 4 of 17 

 

Breath  Sampler.  ReCIVA-RD  Breath  Sampler  does  not  monitor  the  health  of  the  study 
subject  during  sampling.  If  there  is  any  doubt  about  the  health  or  wellbeing  of  the  study 
subject then the breath collection should be stopped immediately.  

We advise monitoring of the breathing frequency of the study subject throughout the breath 
collection. 

 

Always observe the following standard safety precautions: 

  Ensure that the system is correctly assembled before use 

  Do not disassemble any of the individual system components 

  Wear  clean  nitrile  examination  gloves  when  handling  ReCIVA  Breath  Sampler  and 

the Breath Biopsy Mask assembly; chemicals from your hands can affect the results 
and may lead to cross contamination. 

  Wash your hands with soap and water 

– 

do not use alcohol based gels

(e.g. Cutan 

foam) as these leave volatile chemicals behind which may affect the breath collection 

  Do not use the mask if the packaging is damaged or the vacuum seal broken. 

  Use  a  new  collection  kit  with  each  study  subject  (this  includes  a  disposable  mask 

assembly  and  sorbent  tube  assembly)  and  do  not  use  the  collection  kit  if  the 
packaging is damaged 

  Do not use the sorbent tube assembly if it is dropped or damaged prior to the breath 

collection commencing 

  It is recommended to only use the ReCIVA-RD with the collection kits supplied by the 

manufacturer. If you wish to use an alternative, please ensure you carry out your own 
risk assessment to prevent potential harm to the study subject 

  Only use ReCIVA-RD Breath Sampler with the USB lead provided 

  The  study  subject  must  be  monitored  at  all  times  when  using  ReCIVA-RD  Breath 

Sampler 

  If the study subject shows any signs of distress remove ReCIVA-RD Breath Sampler 

and allow them to recover 

  Do not allow the study subject to breath into ReCIVA-RD Breath Sampler without the 

mask and sorbent tube assemblies fitted 

  Do not commence the breath collection until the headset is attached to the patient 

  Do not block the exhaust ports in the base of ReCIVA-RD Breath Sampler 

 

Always  ensure  the  ReCIVA  is  used  with  the  head  strap  to  ensure  a  valid  sample 
collection

.

 

I

f  the  study  subject  wishes  to  hold  the  ReCIVA-RD  Breath  Sampler  as 

well, ensure that they hold it by the sides and not with their fingers or hand over the 
base of the unit as this could block the exhaust ports 

  While  the  ReCIVA-RD  Breath  Sampler  can  be  wiped  down  with  standard 

chlorhexidine cleaning wipes, it must be left for at least an hour between wiping down 
and being used on  the  next study subject  to allow any residual cleaning solution to 
evaporate.  Always  clean  down  the  Breath  Sampler  at  the  end  of  each  sampling 
session 

  Run ReCIVA-RD Breath Sampler on a USB3 port on the laptop/PC. It will not run on 

a USB1 or USB2 port as these do not provide sufficient electrical current. 

  The ReCIVA-RD Breath Sampler is not to be used on children who are too young to 

understand  and  act  on  instruction  from  the  operator  or  on  children  too  young  to 

Содержание Breath Biospy ReCIVA-RD Breath Sampler

Страница 1: ...Ltd 2018 90 0796 Issue 2 OCT 2018 Page 1 of 17 OWLSTONE MEDICAL LTD 183 Cambridge Science Park Cambridge CB4 0GJ UK T 44 1223 428200 E support owlstone zendesk com owlstonemedical com ReCIVA RD Breath Sampler User Manual ...

Страница 2: ...ing up ReCIVA RD Breath Sampler 5 4 0 ReCIVA Software 6 5 0 Taking a Breath Sample 9 6 0 After the breath collection 12 7 0 Servicing 12 8 0 Troubleshooting 13 9 0 Recycling and Disposal 14 10 0 Table of Symbols 14 11 0 Contacts and Support 14 12 0 Appendix 1 Standard Settings for ReCIVA RD Breath Sampler 15 Appendix 2 Adjusting the breath collection parameters 15 ...

Страница 3: ...and operation of this product before using it Using the ReCIVA RD Breath Sampler system in a way that is not specified in this manual could be harmful to health of the operator and co workers This symbol is used to highlight a section explaining particularly important safety considerations Technical Specifications 2 0 Operating Condition Description Input Voltage 5 V Power consumption 4 5W USB 3 0...

Страница 4: ...you wish to use an alternative please ensure you carry out your own risk assessment to prevent potential harm to the study subject Only use ReCIVA RD Breath Sampler with the USB lead provided The study subject must be monitored at all times when using ReCIVA RD Breath Sampler If the study subject shows any signs of distress remove ReCIVA RD Breath Sampler and allow them to recover Do not allow the...

Страница 5: ...CIVA software 02 1898 Sorbent Tube End Cap Jig 90 0796 ReCIVA RD Breath Sampler Instruction Manual Ensure that the system is correctly assembled before use Wear gloves when handling ReCIVA RD The following items supplied by Owlstone Medical Ltd are also required in order to use the ReCIVA RD Breath Sampler 00 0141 CASPER RD Portable Air Supply 00 0137 Breath Biopsy Support Pack Plus one of the fol...

Страница 6: ...sted operation requires elevation 2 Go to the HDF5 installer folder and run the HDF5 installer This is used to write the ReCIVA Breath Sampler log files in HDF5 hierarchical data format 3 Go to the Breath Sampler Controller folder select the Volume subfolder and run setup exe 4 If you want to view the HDF log files install an HDF viewer for example from https www hdfgroup org products hdf5_tools t...

Страница 7: ...f required and additional valves can be ordered from Owlstone Medical Ltd part number 50 1229 The valves are shipped in cardboard tube packaging which needs to be disposed of before the valve is fitted When inserting the valve into the ReCIVA RD ensure the clear film is on the outside so that it can open outwards 5 2 Mask Assembly and Sorbent Tube Assembly The mask and sorbent tube assemblies are ...

Страница 8: ...ing whilst holding ReCIVA RD to their face see Section 6 0 With the secondary Velcro strap adjusters on top of the subject s head attach the hook on one side of the head strap to the attachment point on the ReCIVA RD for most people the hook needs to be in the uppermost position Taking the strap around the study subject s head hook it to the corresponding attachment point on the other side of the ...

Страница 9: ...re 5 Software close button If the ReCIVA RD Breath Sampler is cold lower than 15 C allow it to run for at least 5 minutes before putting it on the study subject This allows the CO2 sensor to warm up preventing condensation which causes it to read full scale 20 2 Give the ReCIVA RD Breath Sampler to the study subject and allow them to breathe through it without putting on the head strap giving them...

Страница 10: ...pattern default 25 seconds learning is shown in the message box Start collecting the study subject s breath collecting is shown in the message box 7 During the breath collection remind the study subject to breathe normally slower comfortable breaths are more effective than rapid breathing patterns Continue to monitor the subject to ensure that they are not distressed If the study subject appears d...

Страница 11: ... on or off light blue on dark blue off The unit also displays the collection time and the study subject s breathing rate 1 Proportion of sample collected 2 Pumps on 3 time taken and breath rate 9 When the collection is complete the message box will show Collect Complete and the collection bar will flash blue and green Remove ReCIVA RD Breath Sampler from the study subject and then close the softwa...

Страница 12: ...For storage return the ReCIVA RD Breath Sampler and the USB lead to its shipping case making sure to put the ReCIVA RD Breath Sampler back in its storage bag ReCIVA RD Breath Sampler must be allowed to dry thoroughly between being wiped down with a chlorhexidine cleaning wipe and being used on another study subject Any fumes from the wipe will invalidate the next study subject collection Servicing...

Страница 13: ...ontroller from the Windows start menu Study subject wants to take a break during sampling This is OK remove the ReCIVA RD Breath Sampler from the study subject let them have a break and then put ReCIVA RD Breath Sampler back on The sampling will pause when ReCIVA RD Breath Sampler is not on the study subject Study subject s face lips or nose are in contact with the sample tubes Adjust the mask unt...

Страница 14: ...roduct For users outside the European Union consult local authorities for correct disposal or contact Owlstone Medical Ltd Table of Symbols 11 0 Description Symbol Description Symbol Caution Consult Accompanying Documents Manufacturer Catalogue Number Date of Manufacture Serial Number Do Not Reuse Do not use if packaging is damaged This mark indicates approval for use in the European Union countri...

Страница 15: ...ronchial enriched Pump B End tidal enriched End bronchial enriched threshold 1 Start bronchial enriched threshold 5 Start end tidal enriched threshold 7 Initial wait before learn 3 seconds Maximum learn time 25 seconds Appendix 2 Adjusting the breath collection parameters ReCIVA RD Breath Sampler has two pumps that can be configured independently Pump A is on the study subject s left and Pump B on...

Страница 16: ...te the collection volume and which portion of the study subject s breath to collect These breath portions are based on the mask pressure and should be used as a systematic guide as to the origin of the breath some mixing within airways will occur Please contact Owlstone Medical if you are in any doubt about the collection settings that you require ...

Страница 17: ...7 Pump Locations Pressing the button bottom left of the screen shows the pump positions as shown below There is also an advanced options screen to provide fine adjustment of the collection parameters Please contact Owlstone Medical before using any of these settings ...