Use mechanical cleaning first to remove particles that may be stuck
to the surface.
To do this, arrange the Travel End Caps in a clean, stainless steel
tray, hole side up.
Bake in a clean oven at 180C for 12-16 hours preferably with carbon
scrubbed or “zero air” flowing through the oven to remove the
contaminants that evolve from the end caps and o-rings.