Pen components
Dose selector
Release button
Internal plunger
Pen cap
Pen body
adaptor
fits onto dose
selector for
easier dialing
Cartridge
Cartridge holder
Pen needle and
protective caps
extension
simply clips onto
release button for
improved grip
Страница 1: ...E USING THE AUTOPEN 24 Failure to follow these instructions may result in an inaccurate dose Data on file Content correct at time of publication Autopen 24 is an easy to use insulin injection pen desi...
Страница 2: ...utton Internal plunger Pen cap Pen body Dose selector adaptor fits onto dose selector for easier dialing Cartridge Cartridge holder Pen needle and protective caps Release button extension simply clips...
Страница 3: ...ternal plunger is extended please refer to step 12 fig 12 to see how to wind it back in Pull off the pen cap and unscrew the pen body from the cartridge holder Load the insulin cartridge into the cart...
Страница 4: ...tective film from a new pen needle Screw the pen needle straight onto the cartridge holder fig 5 Remove the outer protective cap and needle cap from the pen needle If the cartridge contains longer act...
Страница 5: ...start line on the dose selector On subsequent occasions you need only dial to 2 units Repeat dialing and releasing 2 units until insulin appears at the needle tip fig 7a 7b The Autopen 24 is now ready...
Страница 6: ...housing C Hold down the release button towards the internal plunger until the starting line on the dose selector returns to the arrow on the body D Dial up 8 units on the dose selector without screwi...
Страница 7: ...number of units on the dose selector For example figures 8a and 8b show the correct position for 20 units of insulin delivery A B 8 Insert the needle using the injection technique recommended by your...
Страница 8: ...needle cap and unscrew the pen needle Always check your pen needle has been removed Replace the pen cap onto the Autopen 24 Dispose of used pen needles carefully in accordance with the advice given b...
Страница 9: ...remove the cartridge Changing the cartridge 11 Rewind the internal plunger by holding the grey lower part of the pen body between the thumb and finger Turn the pen body until the internal plunger is...
Страница 10: ...xact priming procedures could result in an inaccurate dose being delivered l The needle should be removed and safely disposed of immediately after each injection If the needle is left on it may become...
Страница 11: ...Returning the device will not affect your statutory rights Please check the contents of the package as listed on the outside of the box The Autopen 24 has been thoroughly tested and meets the dose acc...