Panoramic image quality check
Warning
X-rays will be emitted during the performance of the following operations. It is
recommended to use the greatest caution and to comply with local safety regulations
and laws.
1.
Switch ON the unit (see paragraph 9.1).
2.
Open QuickVision software and open the patient "Quality Test". If not present, create a new
patient (Name: "Quality"; Family name: "Test").
3.
Select the "Mouth" icon.
4.
From the "ACQ" toolbar, select the GUI icon to open the virtual keyboard.