GENERAL INSTRUCTIONS FOR USE
9.1 Switching the device ON and OFF
Warning
The unit must be connected to a differential magneto-thermal switch to divide the unit
from the supply. This switch must comply the electrical regulations in force in the
country of installation.
Minimal characteristics at 230V: working voltage 250V, current 10A and differential
current 30 mA.
Minimal characteristics at 115V: working voltage 150V, current 25A and differential
current 30 mA.
Press the power switch located on the upper part of the equipment on the operator side to position
"1".
This will start the "CHECK" function, which is indicated by LEDs lighting up on the equipment
keyboard.
When the "CHECK" function is complete, the green LED (3 - Figure 17) on the equipment
keyboard starts blinking.
To carry out an exam (see also chapter 10):
1. Press ">O<" button on the keyboard to run the equipment axis zero.
2. Run the "GUI" on the PC and wait for the connection between the PC and the equipment; this
status is indicated by the blue LED on the equipment (5 - Figure 17) and on the GUI lighting
up.
3. Select the exam and parameters on the GUI.
4. Press the button ">O<" on the keyboard to run the chin rest automatic positioning accordingly
with the current exam selection. The white LED (8 - Figure 17) will starts blinking (system
waiting for a user's input).
Note
At this moment is no longer possible to change the exam type through a direct selection
on the GUI. In order to undo the current selection:
•
Press the button ">O<" two times to reset the current chin rest position.
•
Make a new exam selection on the GUI.
5. Place the proper chin rest relating to the current exam selection. If the chin rest is correct, the
white LED will become steady ON, otherwise it will blink quickly.
6. Position the patient.
7. Press ">O<" button to set up the equipment in the start exam position; the green LED lights
up.