3.1.1
Precautions while using laser centring devices
For patient positioning, I-Max 3D uses two laser diodes with optical power on the working surface
< 1 mW.
The directive CEI-EN 60825-1 defines the laser as "any device that produces or amplifies
electromagnetic radiation in a coherent manner which includes a wave lengths from 180 nm to 1
mm by means of a stimulated emission". In reference to this directive, the lasers present on the
I-Max 3D are parts of class 1.
A warning label (See picture below) is affixed to I-Max 3D to indicate a laser in class 1 is mounted
internally and caution is advised.
Warning
•
Always keep the room well lit.
•
Do not look into the output windows of laser centring units.
•
Do not stare at the reflections of laser pointers.
•
Instruct the patient to keep his/her eyes closed as long as the laser pointers are
active.
•
Before starting an exam, the patient must remove earrings, glasses, necklaces and
any other item that could reflect the laser beam or be impressed on the
radiographic image.
•
Do not clean the openings of laser centring devices with tools that could modify the
optics. Any cleaning must only be performed by authorized technicians.
•
Operations other than those indicated could cause the emission of dangerous non-
ionizing radiation.