Do  not  resterilize  any  components  of  the  Ovation  Abdominal  Stent  Graft 
System. 

 

Systemic anticoagulation should be used during the implantation procedure 
based  on  hospital  and  physician  preferred  protocol.  If  heparin  is 
contraindicated, an alternative anticoagulant should be considered. 

 

Do  not  bend  or  kink  the  Ovation  Abdominal  Stent  Graft  System  because  it 
may damage the device and/or its components. 

 

Always  use  fluoroscopic  guidance  to  advance  the  delivery  system  and  to 
monitor the implant procedure and the device deployment. 

 

Inaccurate  placement  or  inadequate  seal  may  result  in  increased  risk  of 
leakage into the aneurysm. 

 

Do not continue advancing any portion of the delivery system if resistance is 
felt  during  advancement  of  procedure  accessories  or  of  stent  graft  system. 
Exercise particular care in areas of stenosis, intravascular thrombosis, or in 
calcified or tortuous vessels. 

 

Unless  medically  indicated,  do  not  deploy  the  stent  graft  components  in  a 
location that will occlude arteries necessary to supply blood flow to organs or 
extremities. 

 

Patients  who  experience  hypersensitivity  reactions  during  the  procedure 
should  be  managed  in  accordance  with  standard  recommendations  for 
treatment of patients with radiocontrast agent allergies (e.g., antihistamines, 
corticosteroids, adrenaline).  

5. 

Adverse Events 

5.1.  Potential Adverse Events 

Adverse events that may occur include but are not limited to: 

 

Acute  and  chronic  renal  failure,  renal  microembolism,  renal  insufficiency, 
renal artery occlusion, contrast toxicity; 

 

Allergic reaction to x–ray dye, anti-platelet therapy, device materials; 

 

Anesthetic complications and subsequent attendant problems (aspiration); 

 

Aneurysm enlargement or rupture; 

 

Blood  or  bleeding  events  such  as  anemia,  gastrointestinal  bleeding, 
retroperitoneal bleeding; 

 

Bowel  events  such  as  bowel  ischemia,  bowel  necrosis,  colon  ischemia, 
paralytic or adynamic ileuses, obstruction, fistulas; 

 

Cardiac  events  and  subsequent  attendant  problems  such  as  congestive 
heart  failure,  volume  overload,  arrhythmias,  myocardial  infarction,  chest 
discomfort  or  angina,  elevations  in  creatinine  phosphokinase  (CPK), 
hypotension, hypertension; 

 

Cerebral  events  (local  or  systemic)  and  subsequent  attendant  problems 
such as change in mental status, cerebrovascular accident (hemorrhagic or 
embolic),  reversible  ischemic  neurologic  deficit,  nerve  injury,  transient 
ischemic attacks, paraplegia, paraparesis, paralysis; 

 

Death; 

 

Device events such as deployment or device malfunction, loss of stent graft 
system component integrity, endograft occlusion, migration, or dislodgement, 
endoleak;  

 

Embolic  and  thrombotic  events  such  as  deep  vein  thrombosis, 
thromboembolism,  microembolism,  thrombophlebitis,  phlebothrombosis,  air 
embolism; 

 

General discomfort related to the procedure; 

 

Generalized  inflammatory  response  that  may  be  associated  with  elevated 
levels of systemic mediators of inflammation, elevated temperature; 

Содержание TriVascular

Страница 1: ...810 0005 01 05 Rev B Instructions for Use 810 0005 01 05 EN 0 05 0 0 05 0...

Страница 2: ...ent Selection and Treatment 9 6 1 Individualization of Treatment 9 6 2 Specific Patient Populations 9 7 Patient Counseling Information 9 8 How Supplied 10 8 1 Ovation Stent Graft Delivery Systems 11 8...

Страница 3: ...of the graft and to provide support into which the iliac limbs are deployed the graft body contains a network of inflatable rings that are filled with a liquid polymer that solidifies during the depl...

Страница 4: ...tion Abdominal Stent Graft System is designed to accommodate various aortic anatomies including a range of proximal and distal aortic neck diameters and aneurysm lengths Refer to Table 1 for patient s...

Страница 5: ...ices and or accessories Non aneurysmal proximal aortic neck with a length of at least 7 mm proximal to the aneurysm with an inner wall diameter of no less than 16 mm and no greater than 30 mm and with...

Страница 6: ...d All patients treated with this device must undergo periodic imaging to evaluate the stent graft aneurysm size aneurysm pulsatility device migration leaks device integrity and occlusion of vessels in...

Страница 7: ...Events Adverse events that may occur include but are not limited to Acute and chronic renal failure renal microembolism renal insufficiency renal artery occlusion contrast toxicity Allergic reaction...

Страница 8: ...sible infection urinary tract systemic or localized endograft Pneumothorax Pulmonary respiratory events and subsequent attendant problems such as pulmonary insufficiency pneumonia respiratory depressi...

Страница 9: ...ions for patient selection include but are not limited to Patient s age and life expectancy Co morbidities e g cardiac pulmonary or renal insufficiency prior to surgery morbid obesity Patient morpholo...

Страница 10: ...delivery system the fill polymer kit and the autoinjector The stent grafts are available in the following sizes and configurations Table 2 Ovation Aortic Body Stent Graft sizes Stent Graft Proximal Di...

Страница 11: ...place For single patient use only Do not reuse reprocess or re sterilize Reuse reprocessing or re sterilization may compromise the structural integrity of the device and or lead to device failure tha...

Страница 12: ...nal techniques and in the use of this device The recommended skill knowledge requirements for physicians using the Ovation Abdominal Stent Graft System are outlined below Patient Selection Knowledge o...

Страница 13: ...uoroscopy capability Digital Subtraction Angiography DSA capability Appropriate personnel protection equipment for fluoroscopy Video recorder Power injector with associated supplies Angiography and ex...

Страница 14: ...ated Heating In non clinical testing the Ovation Abdominal Stent Graft System produced the following temperature rises during MRI performed for 15 min of scanning i e per pulse sequence in 1 5 Tesla 6...

Страница 15: ...within the vasculature 10 2 Implant Procedure Do not kink the delivery catheters Doing so may cause damage to the delivery catheters and the TriVascular Ovation Abdominal Stent Graft System Systemic a...

Страница 16: ...ker is toward patient s ipsilateral side CAUTION Rotate entire delivery system as a unit Do not independently rotate catheter sheath or handle 6 Under fluoroscopic guidance retract delivery system out...

Страница 17: ...ill polymer kit Fill Polymer Injection WARNING DO NOT push or pull the delivery system after complete deployment of the proximal stent to avoid inadvertent disconnection of the polymer fill connector...

Страница 18: ...ng of the aortic body most proximal ring 7 Confirm distal iliac limb radiopaque markers are at the appropriate location and that the iliac limb is in the contralateral lumen 8 Retract sheath to deploy...

Страница 19: ...labeling Prepare balloon catheters and other adjunctive devices to be used according to the manufacturer s Instructions For Use Iliac limb aortic body junction The junction may be ballooned using a 12...

Страница 20: ...b Distal Diameter 1 0 X X X 1 2 X X X 1 4 X X X 1 6 X X X 1 8 X X 2 2 X 20 mm Maximum allowable extension 3 To use an iliac limb as an extension refer to the table below Based on the iliac limb distal...

Страница 21: ...paration and use of balloons 13 Re insert angiographic catheter and advance to the suprarenal aorta Perform deployment completion angiography as described above 11 Follow up Imaging Recommendations Tr...

Страница 22: ...n 2 cm above celiac artery origin Inferior Extent AAA Pre op Lesser trochanter of femurs to include femoral bifurcations Post op At least 2 cm distal to the lowest hypogastric artery origin Pre op Les...

Страница 23: ...mperature for excursions during transit only Do not reuse Do not resterilize Keep dry Do not use if package is damaged Sterilized using ethylene oxide STERILE R STERILE R Sterilized using irradiation...

Страница 24: ...ckway Blvd Santa Rosa CA 95403 USA 1 707 543 8800 Authorized Representative MediTech Strategic Consultants B V Maastrichterlaan 127 129 6291 EN Vaals NL T 31 43 306 3320 F 31 43 306 3338 2012 TriVascu...

Отзывы: