3.6.10 Output Level in an Error (Error indicator)
①
Click on "Setup (S)" at top-level menu of the screen, select "Error output" from "Conf. output" drop-
down list, and click on again.
②
A message box as shown in Fig. 49 appears. Output levels in an error are set up at this window.
③
As shown in Fig. 50, select the output level from the drop-down list by clicking on the arrow mark at
right.
④
Given in the table below are output levels.
NOTES :
"Error indicator" does not function in all the errors. For applicable error items, refer to "3.15
Error Messages"
.
⑤
After filling in all the fields required, click on "OK" button. A message box as shown in Fig. 51 then
appears.
Clicking on "OK" at this window changes the previous settings to the new settings just entered.
However, the flowmeter output also changes with the changes in settings made. For safety's sake,
therefore, in applications where the flowmeter output controls a valve or other devices, it is necessary
that the control loop be switched to manual control to ensure that the control loop is unaffected by the
flowmeter output.
Output Level
Analog Output
Pulse Output
Downscale
2.4mA
0Hz
Zero (4mA)
4mA
0Hz
Hold
Holds the last measured reading.
Holds the last measured reading.
Upscale
2.6mA
kHz