Otto Bock TLSO 50R152 Series Скачать руководство пользователя страница 2

50R152=* Ottobock TLSO 

Thank you for using an Ottobock product. The following information provides 

information about the fit and use of the device. Please read the entire document 

carefully before use and ask questions for clarification. 
Practitioner fitting instructions are provided on page 3.

Indications for Use

•  Controls unwanted thoracic flexion
•  Offers post-operative support of the thoracic spine. 

Contraindications for Use

Device should only be applied as instructed by a physician or practitioner.  This device 

is not indicated for use with pregnancy, compromised circulation and/or pulmonary 

function, or any other cardiovascular or skeletal condition which could be compromised 

with the use of the device.Practitioner Fitting Instructions 

CAUTION

Usage Restrictions 

This device is intended for single patient-use only. Reuse of the device is prohibited. 

The daily use of the device depends on the medical indication and the prescribing 

physician orders.

•  Carefully follow all instructions provided by the prescribing physician.

How to Fit and Apply

1.  Position posterior component and loosen hook-and-loop fasteners so the side 

panels wrap around and connect to the anterior/front component. Anterior/front 

bottom edge should be at the symphysis pubis.

2.  Loosen the shoulder straps all the way out and connect to the front loop. Tighten 

as needed. When adjusted the patient can either keep the straps connected and 

slide the TLSO brace over the head to apply OR remove or unattach one side and 

apply like a jacket or back pack by placing one arm through then connecting the 

posterior/back and the anterior/front.

3.  Adjust the anterior thoracic extension so the top edge of the sterna pad is about 1˝ 

below the sternal notch. Check position while patient is sitting for optimal length 

without impingement.

4.  Once the TLSO brace is properly positioned compression can be applied. To begin, 

grasp each of the two handles and simultaneously pull both straight out in front 

of the torso. Once the desired compression is achieved, fasten the pull handles to 

the anterior/front panel. To avoid tangling the lacer cords, position one pull handle 

higher than the other. You can loosen the amount of compression while sitting by 

lifting the handles and then reattach. Tighten again when standing.

5.  To adjust the length, have the patient sit and adjust appropriately with pull tab at 

sternal thoracic extension.

How to Remove

 1.  Instruct patient to remove and apply the TLSO brace by releasing the compression 

handles and removing one of the shoulder straps. Release one side of the side 

panel from anterior/front panel and slide the TLSO brace off the other shoulder like 

removing a jacket or back pack.

Cleaning and Care

•  It is recommended that a snug fitting undershirt be worn between the skin and the 

TLSO Brace.

•  Plastic and metal components and waffled liner may be wiped clean with a damp 

cloth and dried with a towel.

• Sternal pad cover may be removed and washed by hand using mild soap or detergent. 

CAUTION

It is important to ensure the device is kept clean and dry. Do NOT use a washing 

machine, dryer, bleach, or harsh chemicals to clean the device. 

1. Hand wash with mild soap and water.
2. Rinse thoroughly and towel or air dry.

CAUTION

Do not place the device near or on a heat source or open flame. DO NOT USE HEAT TO 

DRY THE DEVICE. 

Practitioner Fitting Instructions 

Measure for the Correct Size

1. Measure the circumference of the patient’s waist.
2. Select appropriate size from chart below.

Adult Sizing Guide

Device Size

Waist Circumference

Back Panel Height

MODEL #

Small

30 – 35˝

17˝

50R152=S

Medium

35 – 40˝

17˝

50R152=M

Large

40 – 45˝

19˝

50R152=L

XL

45 – 50˝

19˝

50R152=XL

Manufacturer’s Limited Warranty

Otto Bock HealthCare, LP (Ottobock) warrants all of its devices, to the original purchaser, 

to be free from defects in materials and workmanship. This warranty applies, subject 

to normal wear and tear, when the devices are used as intended, without unapproved 

modifications, following all Ottobock instructions and requirements; and when they are 

fitted by or under the direct supervision of certified/licensed practitioners. This Limited 

Warranty does not cover device damage caused by accidents, neglect, misuse or operation 

beyond capacity, parts damaged by improper installation, substitution of parts not 

approved by Ottobock, or any alteration or repair by others that, in Ottobock’s judgment, 

materially or adversely affect the device.
The duration of this Limited Warranty varies by product types and is effective from the 

date of delivery to the end-user. Please refer to www.ottobockus.com or call 800 328 4058 

for questions. Ottobock’s sole obligation under this Limited Warranty shall be to repair, 

replace, refabricate the device at no charge, or refund the cost of the device to the original 

purchaser, at Ottobock’s sole discretion. 
THE EXPRESS WARRANTIES SET FORTH ABOVE ARE IN LIEU OF ALL OTHER 