Body weight
Functional ring reference no.
[kg]
[lbs]
4X147=0
40 to 47
88 to 103
4X147=1
48 to 55
104 to 121
4X147=2
56 to 65
122 to 143
4X147=3
66 to 75
144 to 165
4X147=4
76 to 87
166 to 192
4X147=5
88 to 100
193 to 220
4X147=6
101 to 112
221 to 247
4X147=7
113 to 125
248 to 276
New functional rings are very stiff. Ottobock recom
mends pre-compression to ensure that they work cor
rectly.
No pre-compression is necessary with the pre-installed
functional ring and with compressed units.
Open the pre-compression device and insert the func
tional ring.
Tighten the screw of the pre-compression device up to
the stop in order to compress the functional ring com
pletely.
Compress the functional ring from at least
3 minutes
to
at most
15 minutes
.
Open the pre-compression device and remove the func
tional ring.
3.6.4 Replacing the functional ring
Objective
1
2
Loosen the tube clamp of the Harmony P3 unit. Remove
the tube adapter with the prosthetic foot.
Unscrew the retaining screw (1) on the distal end of the
Harmony P3 unit.
Pull the base (2) down distally.
Remove the current functional ring.
Put the new, pre-compressed functional ring on. The
structure of the functional ring only allows attachment in
the correct alignment.
Put the base on.
Secure the retaining screw with Loctite® 241 (636K13)
and screw it in (tightening torque:
7 Nm)
.
Reattach the tube adapter with the prosthetic foot.
18 | Ottobock
Procedure
Harmony System