101
Error while charging the product
LED on
power
supply
Status
LED on
battery
charger
Error
Resolution
Country-specific plug adapter not
fully engaged on power supply
Check whether the country-specif
ic plug adapter is fully engaged on
the power supply.
Non-functional socket
Check socket with another electric
appliance.
Defective power supply
The battery charger and power
supply must be inspected by an
authorised Ottobock Service
Centre.
No connection between battery
charger and power supply
Check whether the charging cable
plug is fully engaged on the battery
charger.
Defective battery charger
The battery charger and power
supply must be inspected by an
authorised Ottobock Service
Centre.
Status
LED
Charging status
indicator (5 LEDs)
Error
Resolution
The LED
ring is lit
in weak
purple
No LED is lit
Distance between bat
tery charger and
receiver of the char
ging unit on the pros
thesis too great. If the
distance is more than
2 mm, the prosthesis
cannot be charged.
Reduce distance
between battery char
ger and receiver of the
charging unit.
LED 2 and 4 are lit up Charger excess tem
perature
LED 1, 3 and 5 are lit
up
Excessively high or
low temperature of the
prosthesis
Check whether the
specified ambient con
ditions for charging the
battery are met (see
Page 95).
The LED
ring is lit
up yellow
LED 3 is lit up
The prosthesis is not
being charged
Distance between bat
tery charger and
receiver of the char
ging unit too great.
Connection may be
improved by reducing
the distance between
the charger and receiv
er of the charging unit.