Turning on the Device

Turn device on via the POWER switch located at the base of the device (Fig.1).
Note: Each time the device is powered up, the extension, flexion, speed and pause 

settings will be the same as when the device was last run. The SET-Lock Switch 

(next to the base power switch) should be in the SET position to make settings.

Start/Stop Control Button

You may stop and restart the CPM at any time by depressing the START/STOP 

button on the Motion Controller. The device will proceed in the opposite direction 

upon restarting.

Danger! 

Adjust Motion Settings only under physician's or Therapist's orders.

Warm-Up Feature

Turn on the device from the POWER Switch located at the base of the device. 

In the motion controller display window you will be prompted with a choice of 

whether to initiate the WARM UP feature or not. You must choose Y (yes) or N 

(no) to proceed.

Selecting the Warm Up Feature:

Choose Y (yes) by pressing the EXT button on the Motion Controller.

By-Passing the Warm Up Feature:

Choose N (no) by pressing the FLEX button on the Motion Controller.

De-selecting the Warm Up Feature:

To turn off the Warm Up feature, turn the device off then on again at the POWER 

Switch. Select N (no) to exit the Warm Up feature prompt.

Setting Range of Motion (ROM) Parameters

The 480 achieves a maximum ROM of -5° extension to 110° flexion. 
The 520 achieves a maximum ROM of -10° extension to 125° flexion. 
The ROM settings are continuously displayed in the Motion Controller.
To set the ROM parameters, press and hold the EXT or the FLEX button while 

simultaneously depressing the desired Up arrow or Down arrow buttons (Fig.2).
The Extension and Flexion angles will change slowly for the initial 5 degrees to 

allow for precise adjustment); following this, the parameters will change rapidly to 

allow for quicker adjustment.