NOTICE
Mechanical stress on the power supply/battery charger
Lack of proper charging functionality due to malfunction.
►
Do not subject the power supply/battery charger to mechanical vibrations or impacts.
►
Check the power supply/battery charger for visible damage before each use.
NOTICE
Operating the power supply unit/charger outside of the permissible temperature range
Lack of proper charging functionality due to malfunction.
►
Only use the power supply unit/charger for charging within the allowable temperature range. The section
"Technical data" contains information on the allowable temperature range (see Page 103).
NOTICE
Independent changes or modifications carried out to the battery charger
Lack of proper charging functionality due to malfunction.
►
Have any changes or modifications carried out only by Ottobock authorised, qualified personnel.
NOTICE
Contact of the battery charger with magnetic data storage devices
Wiping of the data storage device.
►
Do not place the battery charger on credit cards, diskettes, audio or video cassettes.
4.6 Information on Alignment/Adjustment
CAUTION
Use of unsuitable prosthesis components
Falling due to unexpected behaviour of the product or breakage of load-bearing components.
►
Use the product only in combination with components listed in the section "Combination possibilities" (see
Page 68).
CAUTION
Improper assembly of the screw connections
Falling due to breakage or loosening of the screw connections.
►
Clean the threads before every installation.
►
Apply the specified tightening torque values for installation (see the section "Technical data").
►
Observe the instructions for securing the screw connections and the use of the correct length.
CAUTION
Incorrectly secured screws
Falling due to breakage of load-bearing components caused by screw connections coming loose.
►
After completing all settings, the set screws in the tube adapter must be secured before they are tightened to
the specified torque (see the section "Technical data" see Page 103).
►
The tube clamp screws must not be secured but only tightened to the specified torque.
CAUTION
Incorrect alignment or assembly
Falling due to damage to the prosthesis components.
►
Observe the alignment and assembly instructions.