2 Intended use

2.1 Area of application

The product was developed for everyday use. In order to prevent injury or damage to the product, do not use it
for sports or other high impact activities, i.e. jumping from a ladder.

2.2 Environmental conditions

Water:

The prosthetic foot is protected against splash water (e.g. stepping into a shal­
low puddle or walking in the rain).

Do not submerge. Submersion can cause permanent damage.

If water penetration is suspected, turn the prosthetic foot off and allow it to
dry completely before turning it on again.

The battery and charger are not protected against water.

Relative humidity:

10% to 90%, non-condensing

Sand/dust:

No contact with sand or dust. Sand and dust can get into the joint and damage
the mechanics. Protect the product against sand and dust in situations where it
may be exposed: walking on the beach, working on a building site, etc. .

Temperature:

Operation: 0 °C to 45 °C (32 °F to 113 °F)
Charging: 5 °C to 40 °C (41 °F to 104 °F)
Storage: -30 °C to 60 °C (-22 °F to 140 °F)

Impacts/vibrations:

Do not subject the product to mechanical vibrations or impacts.

Electric/magnetic power:

Do  not  use  the  product  in  environments  with  high  levels  of  electric/magnetic
power (e.g. electricity generators, transformers, high-performance radio trans­
mitters, magnetic high-performance transmitters).

2.3 Service life

Prosthetic foot

Expected lifetime given compliance with maintenance intervals: 

6 years

Battery

The manufacturer has determined a maximal service life of one year for the product.

Footshell, protective sock

The product is a wear part, which means it is subject to normal wear and tear.

3 Safety

3.1 Explanation of warning symbols

WARNING

Warning regarding possible serious risks of accident or injury.

CAUTION

Warning regarding possible risks of accident or injury.

NOTICE

Warning regarding possible technical damage.

3.2 General safety instructions

WARNING

Operation of motor vehicles

Risk of accidents due to restricted body function

Observe the applicable legal and insurance regulations for the operation of motor vehicles and have your driv­
ing ability examined and certified by an authorised agency.

CAUTION

Unintentional activation of the relief function

Faulty operation of appliances/machines due to unintentional lowering of the prosthetic foot

Turn the prosthetic foot off and bring it into a neutral position prior to operating control elements (e.g. vehicle
pedals) in a sitting position.

3

1A1-2 Empower

Intended use

Содержание 1A1-2 Empower

Страница 1: ...Instructions for use user 2 1A1 2 Empower...

Страница 2: ...2 Empower is an electronically controlled prosthetic foot with an actively driven ankle joint The prosthetic foot simulates the function of the calf musculature and Achilles tendon for the toe off at...

Страница 3: ...uct in environments with high levels of electric magnetic power e g electricity generators transformers high performance radio trans mitters magnetic high performance transmitters 2 3 Service life Pro...

Страница 4: ...of functionality during use in this section Take any necessary measures e g repair replacement inspection by the manufacturer s customer service etc CAUTION Reaching into the area of the joint mechani...

Страница 5: ...he battery is fully charged the battery life during normal use is up to 8 h Ottobock recommends charging the battery daily Charging the battery before it is drained has a positive impact on its servic...

Страница 6: ...every five seconds Indicates a battery fault Take the battery out of the battery charger and do not use it again Contact your O P professional LED to indicate the charging status on the battery charge...

Страница 7: ...mbling Always use the handrail when walking on stairs Note the product specific information for walking on stairs Walking on stairs with the Empower should be practised The toe off provides support wh...

Страница 8: ...The conformity assessment was drawn up by the manufacturer in accordance with Annex III of the directive The product meets the requirements of the RoHS Directive 2011 65 EU on the restriction of the...

Страница 9: ...t Class II Electrical device Please note the instructions for use Polarity Humidity limits for storage transport Temperature limits for storage transport IP Ingress protection classification 9 1A1 2 E...

Страница 10: ...10 1A1 2 Empower...

Страница 11: ...11 1A1 2 Empower...

Страница 12: ...ck SE Co KGaA Max N der Stra e 15 37115 Duderstadt Germany T 49 5527 848 0 F 49 5527 848 3360 healthcare ottobock de www ottobock com This product and the use thereof may be covered by one or more of...

Отзывы: