Switching the MyModes and changing setting parameters using the Cockpit app is not possible during the
charging process.
•
After disconnecting the battery charger, the orthosis is in the same state it was in before connecting the battery
charger. For example, if the orthosis was switched off before connecting the battery charger, it will also be
switched off after disconnecting the battery charger.
6.1 Connecting the power supply and battery charger
1
2
3
1) Slide the country-specific plug adapter onto the power supply until it locks into place (see fig. 1).
2) Connect the round,
four-pin
plug of the charging cable to the
OUT
receptacle on the battery charger so that
the plug locks into place (see fig. 2).
INFORMATION: Ensure correct polarity (guide lug). Do not use force when connecting the cable plug
to the battery charger.
3) Connect the round,
three-pin
plug of the power supply to the
12 V
receptacle on the battery charger so that
the plug locks into place (see fig. 2).
INFORMATION: Ensure correct polarity (guide lug). Do not use force when connecting the cable plug
to the battery charger.
4) Plug the power supply unit into the wall socket.
→
The green LED on the back of the power supply (see fig. 3) and the green LED on the battery charger light
up.
→
If the green LED on the power supply and the green LED on the battery charger do not light up, there is an
error (see Page 111).
6.2 Connect battery charger to the product
1) Open the charging receptacle cover.
2) Connect the charging plug to the charging receptacle of the product.
INFORMATION: Take note of the plug direction!
→
After the displays are tested, a short beep is emitted followed by three short
vibration signals.
→
If the
symbol lights up, this means an error was identified during the self-
test (see Page 107).
3) The charging process starts.
→
To check the charge level while the battery charger is connected, briefly
press the
button on the control panel.
4) Disconnect the product after the charging process is complete.
→
After disconnecting the battery charger, a vibration signal is emitted followed
by a short beep, and the current status is displayed for approx. 5 seconds
(see Page 107).
→
When the symbol in the
button is lit up green
, the product is switched
on and ready for operation.
5) Close the charging receptacle cover.
INFORMATION
After disconnecting the battery charger, the orthosis is in the same state it was in before connecting the battery
charger. For example, if the orthosis was switched off before connecting the battery charger, it will also be
switched off after disconnecting the battery charger. A descending sequence of beeps is emitted when the bat
tery charger is disconnected while the orthosis is switched off
.
77