If you have changed the settings and return to the Smart DustIQ Explorer you have to change the settings to the ones used
by your instrument.
Click
Confirm
to save and use the settings.
Main window - connections
Once the Dust Explorer software can communicate with the attached DustIQ, the following window appears:
•
The program starts in Normal Network Use, meaning you can have several instruments attached to monitor
•
Changing setting is best done with only one instrument attached
•
Finding an instrument with unkown communication settings is only possible with one instrument attached
•
Allow 5 to 10 minutes for the instrument to stabilise before reading the measurements
•
When started the first time and with a DustIQ straight out of the box: 19.200, 8E1 are the coms settings used and
working.
On the initial 0386900 model, Power Volt is not reported correctly and always 3.2V. Later models (0386910 & 15) report the
actual supply voltage.