Settings in other tools

The following controls are set automatically when REM AutoFit with VerifitLINK 
is open and restored to their original states afterwards:

Adaptation Manager

Step 3

Noise reduction

Current setting in OpenSound Navigator™

Feedback Analyzer

On

Directionality setting

Pinna Omni. For styles with a single 
microphone, directionality is not applicable.

Speech Rescue

Off

Step-by-step Guide

On the Fitting step in Genie 2, enter the REM tool from the More Tools menu 
in the left taskpane. Launch REM AutoFit with VerifitLINK by clicking the 
VerifitLINK button at the bottom of the REM tool.

Step 1: Setup

 

1. 

Select a Measurement method from the drop-down list:  
On ear or Test box 

2. 

Select a Fitting rationale. You can use all available fitting rationales  
on the Verifit, except Camfit.

Содержание REM AutoFit

Страница 1: ...REM AutoFit with Verifit LINK Quick Guide...

Страница 2: ...or Verifit2 4 18 or later is required for simultaneous binaural measurements You must have a network connection between the Verifit and your computer If you are already using the Verifit NOAH module y...

Страница 3: ...ni For styles with a single microphone directionality is not applicable Speech Rescue Off Step by step Guide On the Fitting step in Genie 2 enter the REM tool from the More Tools menu in the left task...

Страница 4: ...D tool from NOAH 5 Prepare for the measurements by following the instructions to the right of the screen These instructions are specific to the measurement method you have selected One of the instruct...

Страница 5: ...m the target match a The automatic measurement process is carried out at 65 dB by default and you can select additional levels 50 dB and 75 dB Gain adjustments are based on the 65 dB measurement b If...

Страница 6: ...on the relevant side to measure one side Click Both to measure both sides binaurally Binaural measurements take place simultaneously if using Genie 2 2019 2 and Verifit2 with software version 4 18 or...

Страница 7: ...ccording to what you select Select Apply all gain adjustments to apply adjustments Select Discard all changes to discard adjustments The data in use when you entered REM AutoFit with VerifitLINK will...

Страница 8: ...21 178600 15500 0577 06 21 oticon com...

Отзывы: