OCT/SLO SERVICE MANUAL Version V.4
21
Procedures Warning
Generally all the following procedures are performed with all 3 parts of the upper covers
removed..
Great care must be undertaken when removing or replacing the covers as
many problems can be started simply by adjustment of the covers
. Specific problems
to look for when replacing the covers are:
•
Pressing on the ‘blue’ fiber may cause degradation of final image quality, or
complete machine failure.
•
Pinching of the electrical cables may cause hindrance of movement of the optical
head or in-ability to close the covers.
** It is the responsibility of the Technical Service Personal to make sure that the
quality of the work and the handling of the system components is done in accordance
to the manufacture requirements and recommendations
.
OTI will not be responsible
for any work or replacement of parts which is performed not to standards or by any
person other than OTI personal.