Össur PROPRIO FOOT Скачать руководство пользователя страница 12

12

•  Press the control button to check the charging level. When the battery is fully charged, all LED lights will turn 

green

Guidelines Regarding Lithium-ion Battery:
•  Avoid frequent full discharges. Recharging a partially charged lithium-ion battery does not cause harm. 
 

Recharging daily is the most efficient routine.

POWER SUPPLY:

Only use the charger supplied.

ATTENTION

: Do not charge when prosthesis is being worn!

NOTE:

  The yellow LED light on the PROPRIO FOOT

®

 will blink when charging. The green will be steady.

•  Do not store the prosthesis/battery in hot places or close to heat-emitting devices like radiators. 

WARNINg SIgNALS FOR POWER LOSS:  (Figure B)

Vibratory and/or auditory warning systems are provided by PROPRIO FOOT

®

 to indicate to the user that power 

loss is forthcoming.  The warning signals are provided in a series of intermittent pulses.  Ensure that the user 
recognizes the warnings.

CAUTION

:  When the PROPRIO FOOT

®

 battery charge is low and the warning pulses for power loss are 

delivered, the foot will return to the latest preset heel height position.  
The foot should not be used when the power is off.

MAINTENANCE

SERVICE

PROPRIO FOOT

®

 is designed and manufactured to provide long and troublefree service intervals. 

It is warranted against defective materials and workmanship for a period of 24 months. To ensure proper
function, it is recommended that PROPRIO FOOT

®

 is inspected every six months by a qualified professional. 

Check for signs of unusual wear.

CLEANINg

Follow these instructions:

•  Remove PROPRIO FOOT

®

 from the cosmetic cover

•  Wipe the foot with a soft cloth moistened with a small amount of isopropanol alcohol.

•  Remove debris or dust without the use of compressed air.

NOTE

: Do not dip the whole foot or pour solvent over the foot. The bearings and seals will be damaged.

Do NOT use compressed air to clean the foot. Air will force pollutants into the bearings which may cause 
malfunctions and early deterioration.

WARRANTY

Össur provides a warranty of 24 months on PROPRIO FOOT

®

.  

See Terms of Warranty for details.

Содержание PROPRIO FOOT

Страница 1: ...Instructions for use Patient EN...

Страница 2: ...Safety Precautions Introduction Technical Specifications Basic Operation Functions Maintentance Warranty 3 5 6 7 8 9 12 12...

Страница 3: ...height 2x Function Press Power ON OFF Hold Sound ON OFF 4x 6x 7x Move to neutral relax Manual motor movement Set heel height 2x Function Press Power ON OFF Hold Sound ON OFF 4x 6x 7x Move to neutral...

Страница 4: ...hazardous metals and should never be disposed of in residential or commercial garbage They should never be incinerated because they may explode Field service should be performed at the recommended in...

Страница 5: ...If un usual movement or product wear is detected in a structural part of a prosthesis at any time the patient should be instructed to immediately discontinue use of the device and consult his her cli...

Страница 6: ...ing phase of level ground walking to improve toe clearance for less stressful and more secure walking When seated PROPRIO FOOT will plantarflex for a more natural appearance of the artificial leg PROP...

Страница 7: ...d electrical products The device has type B applied parts AUTHORIZED REPRESENTATIVE SSUR HF Grjothals 5 110 Reykjavik Iceland 354 515 1300 Physical Properties Weight limit 125kg 275lbs Frame construct...

Страница 8: ...facturer settings will be enabled Your prosthetist may alter the settings to adjust the response Power OFF Figure D To enable the keypad hold and press for at least one second To power OFF hold and pr...

Страница 9: ...s with a beep and vibration Heel height adjustment will start The system beeps and vibrates a second time Immediately lift the prosthesis from the ground to allow for ankle adaptation The actuator wil...

Страница 10: ...of the stair case When reaching the bottom of the steps it will take one stride for PROPRIO FOOT to adapt to the neutral level ground You may also shift weight from sound limb to prosthetic limb to ha...

Страница 11: ...There is no confirmation when the audible feedback is disabled Driving Figure D Caution should be taken with respect to use of PROPRIO FOOT inside of a motor vehicle Inadvertent movement of the ankle...

Страница 12: ...RIO FOOT battery charge is low and the warning pulses for power loss are delivered the foot will return to the latest preset heel height position The foot should not be used when the power is off Main...

Страница 13: ...ssur com ssur Iberia S A Parque Europolis Calle Edimburgo n 14 28232 Las Rozas Madrid Spain Tel 34 91 636 06 93 Fax 34 91 637 21 43 ortesica ossur com ssur Nordic P O Box 67 751 03 Uppsala Sweden Tel...

Отзывы: