24
Problem
Possible cause
Action
blown, replace with a
suitably rated 1A slow blow
fuse
d.)
If all the above causes have
been considered, there may
be a fault on the board.
Please contact Ossila for
information
Power but timer
does not
operate
a.)
The tray door has not
been closed properly
‘TRAY OPEN’ will display
a.)
Ensure the tray door is fully
closed. The drawer should
hold firmly in position when
this is the case.
Continuous
buzzer
a.)
Error in temperature
reading UV Ozone
Cleaner temperature is
extremely high
a.)
Check the temperature
reading by entering the
settings mode (referred to
Section 8.4). If the
temperature is extremely
high, turn the unit OFF and
allow sufficient time for it to
cool down