Optical Scientific Inc
OFS 2000FW User’s Guide
P/N 1910-901-1
Rev. 03/28/13
47
8.2.2 Display and Keypad
The Control Box features a digital display module and 4-position keypad that may be used to enter all user
parameters. To access the Menu press and hold the SET key for at least 3 seconds. This will bring up the first
Menu item which is “Units of Measure”. The following keys are used:
The NEXT (
) key is used to accept the selection and move to the next available parameter.
The SET (
) key is used to accept the selection and exit to operational mode.
The Press the UP (
) or DOWN (
) keys to move up and down to the available choices of a given
parameter such as 9600 baud, 14,400 baud, etc.
The tables describe each user - selectable parameter, available choices, and how to use the keypad to
navigate within the parameter.
Units of Measure
Press
or
keys to select a new unit of measure. Press NEXT to select and
advance to the next parameter or SET to select and exit set-up function.
Meters per Second
Default
Km per Hour
Miles per Hour
Feet per Second
Feet per minute
Averaging Time
Press
or
keys to select a new averaging time. Press NEXT to select and
advance to the next parameter or SET to select and exit set-up function.
3 Seconds
Not recommended for operation. Setup, testing and diagnostics only.
10 Seconds
30 Seconds
60 Seconds
Default
2 Minutes
5 Minutes
10 Minutes
NOTE:
While setting up the meter for the first time or doing trouble shooting, the averaging time should be set to
a short time duration (10 seconds or 3 seconds) for rapid response to any parameter changes. The averaging
time is equivalent to a radio circuit time constant. For any step change in the input, it will take approximately 3 to 5
time constants to reach a 90% Full Scale response. Once the meter is properly set up you can change the
Averaging time to a longer setting such as 60 seconds, to smooth out the data. Regardless of the averaging time,
the display refreshes every 3 seconds.
Baud Rate
RS-232 output only
Press
or
keys to select a new baud rate. Press NEXT to select and advance
to the next parameter or SET to select and exit set-up function.
300 Baud
1200 Baud
2400 Baud
4800 Baud
9600 Baud
Default
14,400 Baud
19,200 Baud
28,800 Baud