![oserio FEG-115 Series Скачать руководство пользователя страница 10](http://html1.mh-extra.com/html/oserio/feg-115-series/feg-115-series_instruction-manual_1655665010.webp)
14
15
Before Using the Monitor
1.6 EMC Guidance and Manufacturer's Declaration
0.01
0.12
0.38
1.2
3.8
12
0.23
0.73
2.3
7.3
23
N/A
N/A
N/A
N/A
N/A
0.1
1
10
100
Recommended separation distance between portable and mobile RF communications equipment and the FLG-351, FEG-113,FEG-123,FEG-115
The FLG-351, FEG-113,FEG-123, FEG-115 is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the FLG-351, FEG-113,FEG-123, FEG-115 can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the FLG-351,
FEG-113,FEG-123, FEG-115 as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
(W)
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where p is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE1:
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE2:
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
Separation distance according to frequency of transmitter
(m)
150 kHz to 80 MHz
d =1.2 P
80 MHz to 800 MHz
d =1.2 P
800 MHz to 2.5 GHz
d =2.3 P
Содержание FEG-115 Series