OSCILLA SM450-FF Скачать руководство пользователя страница 7

 

 

ID: 993 / ver. 206 
 

 

 

6  Use of the program with SM930 and SM950 

 
Basic operation of the program: 
Start  the  program  and  connect  the 

audiometer  to  the  USB  port.  The 

program  will  start  at  125  Hz,  left 

channel  and  80  dB  calibration  level, 

and the tone output will be turned on.  

 

The available buttons that appears in 

the  program  depends  of  which  

audiometer is connected to the PC. 

 

 

SM930

:  

Air 

SM950

Air, Masking, Bone 

 

 

 

The values in the write fields show the difference in dB form Normal calibration. In the right side you will find 

the RETSPL values for each frequency. See section 6. Hearing level offsets for more information. 

 

The  program  can  be  controlled  by  the  mouse,  dragging  the  sliders  and  clicking  the  buttons.  However,  to 

make the calibration process more convenient, it can also be controlled by the keyboard.  

 

  

 

 

 

 

 

 

 

 

 

 

 

 

Changes the frequency. 
 
 
 
 
Move calibration sliders 1 step down, -0.1 dB 
 
 
 
 
Move calibration sliders 1 step up, +0.1 dB 
 
 
 
 
Move calibration sliders 10 step down, -1 dB 
 
 
 
 
Move calibration sliders 10 step up, +1 dB 
 
 
 
 
Change function between air, left, right, and 
bone 

The  program  automatically 

detects  the  audiometer  type, 

and  the  bone  and  masking 

buttons in the top of the main 

screen  and  will  only  be 

present  if  the  audiometer 

supports  bone  and  masking. 

By  using  the  keyboard,  you 

can  keep  your  eyes  on  the 

calibration  equipment  all  the 

time,  instead  of  having  to 

look  at  the  PC  monitor  to 

follow the mouse pointer.  
 

Содержание SM450-FF

Страница 1: ...ID 993 ver 206 1 USB Calibration Software Instructions for Use Oscilla USB100 series USB300 series SM450 FF SM930 SM950 TSM400 TSM500 T830 and T840...

Страница 2: ...th SM930 and SM950 7 6 1 Air conductor calibration 8 6 2 Bone conductor calibration 8 7 Use of the program with TSM400 TSM500 T830 T840 9 7 1 Calibration of probe tone 10 7 2 Pressure calibration 11 7...

Страница 3: ...on file from the computer a USB flash stick or a CD drive and start the installation file Follow the installation windows on the screen and press Finish on the last window The software is now installe...

Страница 4: ...ng Free field Phone2 Insert The program can be controlled by the mouse dragging the sliders and clicking the buttons However to make the calibration process more convenient it can also be controlled b...

Страница 5: ...he left channel place the right telephone on the artificial ear select the right channel and repeat the tone and masking calibration Before you start the calibration be sure the transducer in the drop...

Страница 6: ...cally be set to maximum If it is the Berringer MS16 the volume must be set to 75 and fixed before doing the free field calibration The speaker angle must also be selected in the dropdown menu 0 degree...

Страница 7: ...ore information The program can be controlled by the mouse dragging the sliders and clicking the buttons However to make the calibration process more convenient it can also be controlled by the keyboa...

Страница 8: ...left channel place the right telephone on the artificial ear select the right channel and repeat the tone and masking calibration Before you start the calibration be sure the transducer in the dropdow...

Страница 9: ...the RETSPL values for each frequency See section 6 Hearing level offsets for more information The program can be controlled by the mouse dragging the sliders and clicking the buttons However to make...

Страница 10: ...on of probe tone To calibrate the probe tone an artificial ear with a 2cc coupler2 is required Place the probe in a 2cc coupler Adjust the sliders so all frequencies from 226 Hz to 4 kHz reach 100 dB...

Страница 11: ...ssure when calibrating The manometer must be able to measure at least 500 daPa to 500 daPa with an accuracy of 0 5 Follow the guide in the calibration software 1 Press Start pressure calibration 2 Be...

Страница 12: ...essure and type in the value in the input field and press Next 5 Read out the negative pressure and type in the value in the input field and press Next 6 The pressure calibration is now completed Unpl...

Страница 13: ...can be done by selecting 500 Hz and then position the telephone to get the biggest readout on the calibration equipment By means of the five calibration sliders calibrate the sound pressure for the fi...

Страница 14: ...D 993 ver 206 14 7 4 Time and date The TSM400 TSM500 T830 and T840 has internal date and time Real Time Clock RTC To set this similar to the time and date from the PC click the button Set Date and Tim...

Страница 15: ...ft channel place the right telephone on the artificial ear select the right channel and repeat the tone and masking calibration Before you start the calibration be sure the transducer in the dropdown...

Страница 16: ...ml file that is placed in the program folder RetSPL values 0 0 to 200 0 RetSPL_TDH39_NameOfStandart ISO 389 1 1998 RetSPL_TDH39_NameOfStandart RetSPL_TDH39_125 45 RetSPL_TDH39_125 RetSPL_TDH39_250 25...

Страница 17: ...ID 993 ver 206 17 ISO RETSPLs ANSI S3 6 2010 RETSPLs...

Страница 18: ...mode all corrections and RETSPL values are included automatically i e 80 SPL in the calibration software results in a read out of 80 SPL on the level meter assuming microphone and mastoid corrections...

Страница 19: ...05 50 95 05 250 25 5 0 00 70 95 50 500 11 5 0 02 80 91 48 750 7 5 0 01 80 87 49 1000 7 0 0 01 80 87 01 1500 6 5 0 04 80 86 54 2000 9 0 0 11 80 89 11 3000 10 0 0 25 80 90 25 4000 9 5 0 40 80 89 90 6000...

Страница 20: ...the dB level meter will also show 105 5dB SPL i e 80dB HL 25 5 dB at 250 Hz 105 5dB SPL Bone condcutor When calibrating an audiometer with B 71 at 50 dB VFL on 1000 Hz from the calibration software th...

Страница 21: ...the calibration is finished If you have a free field audiometer and have done the Free field calibration too remember to check mark the Free field calibrated In the settings window you are able to set...

Страница 22: ...ivity force in mV N 120 mV N to 140 mV N combined with small frequency corrections will be stated on the calibration certificate when you get a factory calibration from e g Br el Kj r for the BK4930 m...

Страница 23: ...uipment information In order to get the type and serial numbers for calibration equipment on the calibration certificate you have to enter this information in the calibration software Click on Setting...

Страница 24: ...The electrical installation to which the device is connected complies with EN IEC requirements The device is used in accordance with the instructions for use The manufacturer reserves the right to wai...

Отзывы: