background image

 

 

 

Learning the System 

Intended use 

The GluNEO Lite Smart test  meter  is used  with  GluNEO Lite Smart test  strip  for the 
measurement of glucose in fresh  whole  blood  samples(Capillary and/or  venous). 
It should  not  be used  for the diagnosis or screening of diabetes or for the testing  of new  born 
babies. 
GluNEO Lite Smart Blood  Glucose Monitoring System  is intended for  use  outside  the  body 
only(in-vitro  diagnostic use). 
GluNEO Lite Smart Blood  Glucose Monitoring System  is for  self-testing by people  with 
diabetes and  by healthcare professionals. 
Recorded results  make  it easier  for your  doctor  to access  your  metabolic control. 
However, the self-testing should  not take the place  of overall  diabetes management 
assessment from  your  doctor. 

 

CAUTION 

- Do  not  use  system  for any  other  purpose than  blood  glucose test. 
- Cholesterol concentrations > 510  mg/dL  or  triglyceride concentrations > 3,400  mg/dL 

may  produce elevated reading. 

- Icodextrin does  not  interfere with  GluNEO Lite Smart test  strips. 
-  Do  not  use  during  or  soon  after  xylose  absorption  testing.    Xylose  in  the  blood  will  cause 

interference. 

- Before  using  any  product, please  read  all instructions. 
-  Consult  with  your  physician  or  diabetic  healthcare  professional  about  daily  manage- ment  

of  your   diabetes  and  proper   use  of  the  glucometer.  If  you  have   any  questions  about  the  
GluNEO Lite Smart  product, please  contact  your  local  representative. 

 

 

WARNING

 

- Before  using   GluNEO Lite Smart system, read  all  instructions in this  manual, and  practice 

for  accurate  and  safe  test.  You  should    have  commentary  recommendation  from  your 
diabetes  care  professional  for  the  proper   use  of  the  meter  and  daily  management  of  your 
diabetes. 

-  If  you  think  your  blood  glucose results   are  too  low  or  too  high,  or  if  the  results   are 

doubtful, please  contact  your  doctor. 

-  If  your  blood  glucose result  is  unusually  low  or  high,  or  you  do  not  feel  the  result  is correct,  

repeat  the  test  again  with  a new  test  strip. 

- If the results  are still inconsistent, please  consult  with  your  physician before  making  any 

decision in controlling your  diabetes. 

Содержание GluNEO Lite Smart

Страница 1: ......

Страница 2: ......

Страница 3: ...sional before making any changes to your diabetes management The GluNEO Lite Smart Blood Glucose Monitoring System is for in vitro diagnostic use only Customer Service is available 24 hours a day 7 da...

Страница 4: ...ose Monitoring System 5 Do not change your medication based on the GluNEO Lite Smart meter test results without contacting your physician or healthcare professional first 6 Do not assemble or disassem...

Страница 5: ...sugar 27 Reviewing your averages 27 Reviewing your results with activated user ID 28 Deleting test results 28 5 Alternate Site Test Lancing and sampling from an alternate site area 29 31 Lancing devi...

Страница 6: ...rin does not interfere with GluNEO Lite Smart test strips Do not use during or soon after xylose absorption testing Xylose in the blood will cause interference Before using any product please read all...

Страница 7: ...strip 3 Reusable Lancing Device 4 Lancets 5 Instructions for use 6 Warranty Registration Card 7 Carrying Case 8 Patient Logbook 9 3V Coin 1EA Battery CR2032 Your GluNEO Lite Smart Blood Glucose Monit...

Страница 8: ...eases and discards test strip Serial Number Battery Cover CAUTION 1 Do not use GluNEO Lite Smart glucose test meter in a dry environment especially If synthetic materials are present Synthetic clothes...

Страница 9: ...ooting Symbol DATE Delete Results Average Test Result System Before Having a Meal After Having a Meal After Taking Medication After Sport Activity Control Solution Symbol Blood Drop Symbol for Test Us...

Страница 10: ...t strips Throw GluNEO Lite Smart test strips and vial away after the expiry date 5 Do not use the test strips after the expiration date printed on the package or vial since it may cause inaccurate res...

Страница 11: ...glucose test strip 3 Practice using the lancing device and become accustomed with its use 4 Lancets and blood glucose test strips should be single use only 5 Wash your hands in warm clean running wat...

Страница 12: ...ing either or buttons to set the year push button User Activity Option After pressing and releasing either or buttons to turn the User Activity Option on off push button Date Time After pressing and r...

Страница 13: ...programming the number of desired alarms 5 alarms possible push button CAUTION 1 VERY IMPORTANT Set the right test unit Your GluNEO Lite Smart Meter is pre set to mg dL as test unit A wrong test unit...

Страница 14: ...code number on the glucose test meter every time Check the code on the vial to see if the codes match each other CAUTION 1 Do not bend the glucose test strip to prevent the automatic code recognition...

Страница 15: ...se test meter or blood glucose test strip is not functioning properly If your blood glucose test results are not consistent with your symptoms or if you think results are not accurate If you have drop...

Страница 16: ...range CAUTION 1 Control solution is for in vitro diagnostic use only This means that it is only used for testing outside the body 2 Control solution is exclusively made for the GluNEO Lite Smart Bloo...

Страница 17: ...Do not use beyond expiration date Once opened the control solution expires after 3 months If your control solution test falls out of range please follow the next steps before contacting customer suppo...

Страница 18: ...ls out of range do not perform a glucose test and please contact your local representative 2 The glucose control solution s measurement range has nothing to do with individual s blood glucose level as...

Страница 19: ...us this also comes with a transparent tip for AST Alternative Site Testing CAUTION 1 Never use a lancet that has been used by someone else Sharing lancet could lead to contamination 2 If the lancing d...

Страница 20: ...ot enough blood gently massage the area until a sufficient sample is collected STORAGE Products must be stored at room temperature protected from sunlight or moisture CLEANING AND DISINFECTION Clean a...

Страница 21: ...test strip code number and adjusts the meter accordingly A blinking test strip will appear at the top of the screen which indicates that the meter is ready for testing CAUTION 1 If you do not see the...

Страница 22: ...nsmit infectious disease even after you have conducted cleaning and disinfection measures 2 Avoid getting hand lotion oil dirt or debris in or on the lancets single use only and the lancing device 3 W...

Страница 23: ...thick or calloused skin 6 Pull the end of the lancet device back 7 To prick your finger push the center button CAUTION 1 If the blood smears or runs do not use such blood sample Dry the area and gentl...

Страница 24: ...f your finger until you get the beep sound The test will automatically start Please respect recommended quantity of blood sample Too much blood may cause malfunction of your meter by contaminating it...

Страница 25: ...your result then press and release the power button If you do not select to use User Activity option during set up press the power button If you selected to use the User ID option during set up press...

Страница 26: ...ous waste in your area Be sure to contact your local representative for proper disposal 1 Push the needle into the protective cover 2 Pull the lancet out and discard accordingly CAUTION 1 Pushing the...

Страница 27: ...er to calculate ReviewingYour blood sugar Press and release button The most recent result appears first Note the time and day of your blood sugar result Press and release the button and your previous...

Страница 28: ...results of the selected user ID are displayed DATE DeletingTest Results Deleting individual test result To delete any individual test result in the memory press or button for 3 seconds Then displayed...

Страница 29: ...used when sugars are stable before meals and before bedtime However when sugars are changing blood from the fingertip may show these changes sooner than blood from other body areas LIMITATION 1 Alter...

Страница 30: ...a of the forearm or upper arm away from bone visible veins and thick hairs of the arm Sometimes there is less blood flow to these areas than to the fingertip To help you get enough drops of blood you...

Страница 31: ...vice with anyone The lancing device clear tip is used for alternate site testing Lance the test site with the lancing device 1 Remove the lancing tip by twisting off 2 Insert lancet and replace with t...

Страница 32: ...ure site each time you test Repeated punctures in the same spot may cause soreness and calluses 3 If bruising occurs at an alternate site or you have difficulty getting a sample consider sampling from...

Страница 33: ...r in 1 bleach 9 water ratio Dampen a cloth with this solution and wipe the lancing device thoroughly Soak only the tip for at least 30 minutes in the disinfectant solution Do not soak the lancing devi...

Страница 34: ...rect sunlight and heat Do not freeze After removing a test strip from the vial immediately close it tightly Record the discard date date vial opened plus 6 months on the test strip label Do not use if...

Страница 35: ...from the date you first open a new vial of test strips Throw GluNEO Lite Smart test strips and vial away after the discard date 5 Do not use the test strips after the expiration date printed on the p...

Страница 36: ...at the same time as buzzer sound the LCD window will display the P and the operating time count by one second units The meter will begin to transfer Bluetooth signal to your mobile phone and to count...

Страница 37: ...est results from the GluNEO Lite Smart meter to a computer You will need to download the software from the following web site http www osanghc com and purchase the computer cable from your local repre...

Страница 38: ...ium battery should be used Follow up Action 1 Check your meter with the glucose control solution 2 Perform your test again 3 If you are experiencing symptoms that are not consistent with your blood gl...

Страница 39: ......

Страница 40: ...lp you identify certain problems but may not solve all the problems that could occur Contact your authorized representative or customer support center of OSANG Healthcare Co Ltd if the problem continu...

Страница 41: ...t a new glucose test strip and perform your test again If Sun is blinking avoid direct sunlight and re test Contact your local representative or customer support at OSANG Healthcare Co Ltd if the prob...

Страница 42: ...Change the batterry and if the problem persists contact your local represen tative or customer support at OSANG Healthcare Co Ltd...

Страница 43: ......

Страница 44: ...connecting device Once the app is ready to receive your measuring value following message is appeared Turn off the power of the device by clicking power button push Bluetooth button according to the m...

Страница 45: ...th GluNEO Lite Smart device If the connection status between app and GluNEO Lite Smart is not set the display of product number registration will be appeared Register product serial number into your A...

Страница 46: ...Display of product number registration In case that registered product number is different from the device product number...

Страница 47: ...ely transferred below display will be appeared Select Select Save button or Save all button and store measuring value into your App If several people use one GluNEO Lite Smart device together select S...

Страница 48: ...Healthcare Co Ltd has no knowledge of the performance of the instrument when the test strip is altered or modified in any manner Purchaser must contact the customer service department of OSANG Health...

Страница 49: ...ever there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to radio or television reception which can be determined by t...

Страница 50: ...20 60 Bluetooth Bluetooth Smart Bluetooth Low Energy...

Страница 51: ......

Страница 52: ...MEMO...

Страница 53: ...132 Anyangcheondong ro Dongan gu Anyang si Gyeonggi do Korea 14040 www osanghc com INFM001LS IMC01538...

Отзывы: