Introduction

Installation and Site Specifications

Part No. 7B6090

Operator’s Guide

Version 2.0

VITROS ECi/ECiQ Immunodiagnostic System

1-9

Environmental Specifications

The environmental limits for normal operation of the system are defined 
below and shown in 

figure 1-2

.

• BTU Output: 4100 BTUs per hour

• Operating Temperature: 15–30°C (59–86°F)

• Site Relative Humidity: 15–75% RH noncondensing

• Altitude: -0.1524/2.439 km (-500/8000 ft.)

Figure 1-2. Environmental Requirements

System Relocation

The system is mounted on casters to facilitate its relocation within the 
laboratory. The new site must meet the same space, electrical, and 
environmental requirements as specified for the original site.

30.0°C
(86°F)

26.0°C
(78°F)

15.0°C
(59°F)

15%

50%

75%

Relative Humidity

Temperature