Introduction
Installation and Site Specifications
Part No. 7B6090
Operator’s Guide
Version 2.0
VITROS ECi/ECiQ Immunodiagnostic System
1-9
Environmental Specifications
The environmental limits for normal operation of the system are defined
below and shown in
.
• BTU Output: 4100 BTUs per hour
• Operating Temperature: 15–30°C (59–86°F)
• Site Relative Humidity: 15–75% RH noncondensing
• Altitude: -0.1524/2.439 km (-500/8000 ft.)
Figure 1-2. Environmental Requirements
System Relocation
The system is mounted on casters to facilitate its relocation within the
laboratory. The new site must meet the same space, electrical, and
environmental requirements as specified for the original site.
30.0°C
(86°F)
26.0°C
(78°F)
15.0°C
(59°F)
15%
50%
75%
Relative Humidity
Temperature