background image

4

Once the input cable installation has been made,
the electronic noise performance of the 142IH can
be predicted by calculating the cable capacity from
the above information and adding the capacity
expected from the detector.

3.2.  CONNECTION TO AMPLIFIERS

Either or both the E and T outputs of the 142IH can
be connected to an amplifier input for further
processing. The output impedance through either of
these connectors is 93

S

, providing a series

termination for 93

S

 cable so that long cable lengths

can be driven easily. Although the outputs are
marked E (for Energy) and T (for Timing), the pulse
characteristics are identical and the circuits are
interchangeable.

In an energy spectroscopy system, the preamplifier
output is furnished into a shaping main amplifier. In
a timing spectrometer system, the preamplifier
output is furnished into a timing filter amplifier. With
the dual output connections on the 142IH, the
signals can be furnished simultaneously to both
types of spectrometer systems. If either the E or T
output connector is not being used, it should simply
be left open-circuited (unterminated).

3.3.  INPUT POWER

Power for the 142IH is supplied through the Power
Cable that is captive through the rear panel of the
unit. The normal connection for this power cable is
included on the rear panel of the mating ORTEC
amplifier, furnishing ±12 V and ±24 V from the bin
and power supply in which the amplifier is operated.
If this facility is not available or if such a connection
would increase the loading on the bin and power
supply beyond its maximum rated capacity, use an
ORTEC 114 Preamplifier Power Supply to furnish
the operating power requirements through the
captive cable. The ORTEC 114 can furnish power
for two ORTEC preamplifiers simultaneously if
desired.

3.4.  TEST PULSE

A voltage test pulse for energy calibration can be
accepted through the Test Pulse input connector on
the 142IH without the use of an external charge

terminator. The test input of the preamplifier has an
input impedance of 93

S

 and its circuitry provides

charge injection to the preamplifier input. The
shape of this pulse should be a fast risetime (less
than 40 ns) followed by a slow exponential decay
back to the baseline (200 to 400 µs). While the test
pulses are being furnished, connect either the
detector (with bias applied) or the equivalent
capacitance (without bias applied) to the Input
connector on the 142IH.

The Test Pulse input may be used in conjunction
with the output of a pulser such as the ORTEC 419
or 448 to calibrate the preamplifier E output
amplitude in terms of energy for calibration of a
multichannel analyzer. However, due to stray
coupling between the test circuit and other portions
of the preamplifier circuitry, the transient
performance of the preamplifier is best determined
by connecting the actual detector signal through the
Input connector instead of using the pulse generator
output signals for this calibration.

A voltage test pulse for transient response in the
142IH can be accepted through a charge terminator
and into the Input connector on the 142IH. If
external capacitance is to be included for these
tests, an SHV tee can be inserted between the Input
connector and the charge terminator, and this will
then accommodate the test capacitances. Do not
furnish any bias during these tests.

3.5.  DETECTOR BIAS INPUT

Operating bias for the detector is supplied to the
Bias connector on the 142IH and, through a filter
and a large bias resistance, to the Input signal
connector. From there it is furnished out through the
signal input cable to the detector.

Connect a cable from the detector bias supply
(ORTEC 428 or 459 is typical) to the Bias connector
on the 142IH. Type SHV connectors are used in this
high-voltage circuit and the mating cable should be
furnished with the bias supply module.

Содержание 142IH

Страница 1: ...Model 142IH Preamplifier Operating and Service Manual Printed in U S A ORTEC Part No 717590 1202 Manual Revision C...

Страница 2: ...n so that a Return Authorization Number can be assigned to the unit Also ORTEC must be informed either in writing by telephone 865 482 4411 or by facsimile transmission 865 483 2133 of the nature of t...

Страница 3: ...TS 3 2 4 CONNECTORS 3 2 5 ELECTRICAL AND MECHANICAL 3 3 INSTALLATION 3 3 1 CONNECTION TO DETECTOR 3 3 2 CONNECTION TO AMPLIFIERS 4 3 3 INPUT POWER 4 3 4 TEST PULSE 4 3 5 DETECTOR BIAS INPUT 4 4 OPERAT...

Страница 4: ...a hazard that could result in bodily harm if the safety instruction is not observed CAUTION Indicates a hazard that could result in property damage if the safety instruction is not observed Please re...

Страница 5: ...he instrument during external cleaning use only enough liquid to dampen the cloth or applicator SAFETY WARNINGS AND CLEANING INSTRUCTIONS Cleaning Instructions To clean the instrument exterior Unplug...

Страница 6: ...vi...

Страница 7: ...ry before connecting a low impedance a cable or any other capacitive device to the Input connector on the preamplifier 2 Discharge the bias circuitry before making any connections to the Input connect...

Страница 8: ...resistance on the input so that relatively long cable lengths can be used between the detector and the preamplifier without disrupting the system stability A Test Pulse connector with built in charge...

Страница 9: ...t polarity is opposite from the input pulse polarity output pulse polarity is the same as bias polarity 2 4 CONNECTORS INPUT and BIAS Type SHV TEST PULSE E and T Type BNC POWER CABLE 10 ft 3 m captive...

Страница 10: ...se for energy calibration can be accepted through the Test Pulse input connector on the 142IH without the use of an external charge terminator The test input of the preamplifier has an input impedance...

Страница 11: ...n ORTEC 114 Preamplifier Power Supply Figure 4 1 is a simplified block diagram of the circuits in the 142IH Preamplifier The complete circuit is shown in schematic 142IH 0201 S1 included at the back o...

Страница 12: ...amplifier Input 4 4 TIMING OUTPUT The T output connector provides an alternate path for the same output pulses that are furnished through the E output connector The intent is to provide both connectio...

Страница 13: ...t shown for zero external capacity In any case the Input connector and capacitors when used should be completely shielded electrically A wrapping of aluminum foil around the Input connector or a shiel...

Страница 14: ...d parameters By use of the above equation and with a knowledge of the noise of the preamplifier the noise width of the detector can be determined The significance of this noise width in evaluating the...

Отзывы: