EMC Declaration (continued)
Guidance and manufacturer’s declaration – electromagnetic immunity
The “Digital Blood Pressure Monitor ORO-N4 CLASSIC” is intended for use in the
electromagnetic environment specified below. The customer or the user of the “Digital
Blood Pressure Monitor ORO-N4 CLASSIC” should ensure that it is used in such an
environment.
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic
environment –
guidance
• Conducted RF
IEC 61000-4-6
• Radiated RF
IEC 61000-4-3
• 3 Vrms
150 kHz
to 80 MHz
• 3V/m
80 MHz to
2,5 GHz
• 3V
• 3V/m
Portable and mobi-
le RF communica-
tions equipment
should be used
no closer to any
part of the “Digital
Blood Pressure
Monitor ORO-N4
CLASSIC”,
including cables,
than the
recommended se-
paration distance
calculated from
the equation
applicable to the
frequency of the
transmitter.
Recommended se-
paration distance
• d=1.2√P
• d=1.2√P
80MHz to
800MHz
• d=2.3 √P
800MHz
to 2.5 GHz