Oroboros Instruments Oxygraph-2k Скачать руководство пользователя страница 6

MiPNet19.01 

B: OroboPOS service

 

24

 

 

 

OROBOROS INSTRUMENTS

 

 

OROBOROS Oxygraph-2k

 

evaporation of ethanol. Before an experiment, the ethanol is siphoned off 

and the chamber is washed with distilled water (

MiPNet06.03

)

 

5.2. Short-term shelf storage

 

 

For  shelf  storage,  unplug  the  POS-Connector  from  the  O2k-Main  Unit. 
Clean the sealing tip and membrane with distilled water. The POS head is 

maintained moistened by applying a cup on  the sensor head to 

prevent 

the drying out of the electrolyte

. Store in the dark. 

 

5.3. Long-term storage

 

 

For storage of the POS for several months, the sealing tip and membrane 
are removed. Wash the electrolyte off the POS with distilled water.  Even if 
the  membrane  is  not  damaged,  remove  it  by  gripping  the  membrane 

holding ring with the groove in the lower end of the membrane ring holder 
of  the  mounting  tool  (Fig.  2).  The  POS  head  is 

stored  dry  and  in  the 

dark

.  Check  for  any  moisture  and  salt  contamination  in  the  electrical 

connector  of  the  POS  head.    In  this  case,  wash  with  distilled  water  and 

subsequently with pure methanol, dry at 60 

o

C for 24 h.   

 

6.

 

Replacement of the OroboPOS head

 

 

After  cleaning  and  drying  the  gold  cathode  contact  or  drying  the  sensor 

body  at  60  °C  for  a  few  days,  the  zero  current  should  be  reduced.  In 
addition,  the  zero  current  of  the  bare  cable,  without  sensor  head 

connected, is tested for any leak currents. If the latter test excludes any 
sources  of  leak  currents  other  than  the  POS,  and  POS  service  is  not 

successful, the sensor head must be replaced. 

 

 

A  new  sensor  head  can  be  screwed  onto  the  sensor  connector,  if  the 

old sensor head has been irreversibly damaged or should be replaced. 

 
 

7.

 

References

 

 

 

Gnaiger  E  (2008)  Polarographic  oxygen  sensors,  the  oxygraph  and 
high-resolution  respirometry  to  assess  mitochondrial  function.  In: 
Mitochondrial  dysfunction  in  drug-induced  toxicity  (Dykens  JA,  Will  Y 
eds) John Wiley: 327-352.

 

»

 

 

Gnaiger  E,  Forstner  H,  eds  (1983)  Polarographic  Oxygen  Sensors. 
Aquatic  and  Physiological  Applications.  Springer,  Berlin,  Heidelberg, 
New York: 370 pp. »

 

 

Hitchman  ML  (1983)  Calibration  and  accuracy  of  polarographic  oxygen 
sensors.  In:  Polarographic  oxygen  sensors,  Gnaiger  E  and  Forstner  H, 
eds., Springer, Berlin, Heidelberg, New York: 18-30. »

 

 

 

Hitchman  ML,  Gnaiger  E  (1983)  A  thermodynamic  consideration  of 
permeability  coefficients  of  membranes.  In:  Polarographic  oxygen 
sensors,  Gnaiger  E  and  Forstner  H,  eds.,  Springer,  Berlin,  Heidelberg, 
New York: 31-36. »

 

 

 

 

Full version » 

www.bioblast.at/index.php/MiPNet19.01B_POS-Service

 

Next step – O2k-Core Manual D

 » 

MiPNet19.01D O2k-Calibration

 

 

Содержание Oxygraph-2k

Страница 1: ...s of operation Performance of the OROBOROS Oxygraph 2k according to instrumental specifications is obtained only with oxygen sensors which are maintained in a good functional state POS service is required if 1 a new sensor is prepared 2 the raw signal at air calibration declines 3 signal noise is high 4 the time response is prolonged time constant 10 s and biphasic 5 the oxygen signal at zero cali...

Страница 2: ...Net14 01 For sensor service remove the black POS seal tip 1 It is normal to see many small bubbles in the electrolyte reservoir This does not indicate that this caused a problem while the sensor was actually in use Remove the PEEK membrane ring 2 3 and membrane 4 Fig 2 Wash off the electrolyte with distilled water For cleaning of the anode and cathode the sensor head is best removed from the Orobo...

Страница 3: ...ig 3 Hold the sensor in a vertical position and polish the cathode for one minute in a figure eight motion Fig 3 Wash the aluminum oxide powder carefully off the sensor with distilled water and wash the cleaning tools You may further clean the gold cathode with ultrasonic treatment at low power for up to 30 seconds while immersing the tip of the sensor head in distilled water In extreme and rare c...

Страница 4: ...rystals Fig 4 As electrolyte use KCl solution 1 mol dm 3 74 56 g potassium chloride per litre in distilled water Add distilled water to the electrolyte powder Fig 1 up to the 10 ml mark Alternatively dissolve 1 49 g KCl in distilled water with a total volume of 20 ml Store at room temperature or 4 C in a closed bottle To prevent the formation of gas bubbles in the electrolyte heat the electrolyte ...

Страница 5: ...2k is switched on Such stabilization may require several hours For this purpose the O2k may be left on overnight 4 Electrical cable connection Connect the POS cable to the O2k Main Unit avoiding bending and torsions of the cable Insert the male plug of the POS cable into the female connector next to the control light of the stirrer The red dot on the male plug has to face accurately upwards when i...

Страница 6: ...t or drying the sensor body at 60 C for a few days the zero current should be reduced In addition the zero current of the bare cable without sensor head connected is tested for any leak currents If the latter test excludes any sources of leak currents other than the POS and POS service is not successful the sensor head must be replaced A new sensor head can be screwed onto the sensor connector if ...

Страница 7: ...ights of the Oxygraph 2k left chamber between the MAINS and STIRRER lights right chamber between the STIRRER and COMM lights The ridge on the male plug has to face accurately downwards when inserting the plug before rotating the connector screw and fixing it finger tight O2k Manual MiPNet19 01A Oxygraph 2k start high resolution respirometry MiPNet14 01 Electrostatic discharge ESD damage and protec...

Отзывы: