background image

the primary mirror is as centered in the secondary mirror as 
possible. It may not be perfectly centered, but that is OK. Now 
tighten the three small alignment screws equally to secure the 
secondary mirror in that position.
If the entire primary mirror reflection is not visible in the sec-
ondary mirror, as in Figure 6c, you will need to adjust the tilt of 
the secondary mirror. This is done by alternately loosening 
one of the three alignment screws while tightening the other 
two, as depicted in Figure 8. The goal is to center the primary 
mirror reflection in the secondary mirror, as in Figure 6d. Don’t 
worry that the reflection of the secondary mirror (the smallest 
circle, with the collimation cap “dot” in the center) is off-center. 
You will fix that in the next step.

adjusting the Primary Mirror

The final adjustment is made to the primary mirror. It will need 
adjustment if, as in Figure 6d, the secondary mirror is cen-
tered under the focuser and the reflection of the primary mir-
ror is centered in the secondary mirror, but the small reflection 
of the secondary mirror (with the “dot” of the collimation cap) 
is off-center.
The tilt of the primary mirror is adjusted using the three sets of 
two collimation screws on the back end of the optical tube. 
Adjusting the tilt of the mirror requires a “push-pull” technique 
involving adjustment of each set of collimation screws. Loosen 
the one of the screws one full turn, and then tighten the adja-
cent screw until it is tight as in Figure 9 (do not overtighten.) 
Look into the focuser and see if the secondary mirror reflec-
tion has moved closer to the center of the primary. You can tell 
this easily with the collimation cap and mirror center mark by 
simply watching to see if the “dot” of the collimation cap is 
moving closer or farther away from the ring on the center of 
the primary mirror. Repeat this process on the other two sets 
of collimation screws, if necessary. It will take a little trial and 
error to get a feel for how to tilt the mirror in this way. When 
you have the dot centered as much as possible in the ring, 

your primary mirror is collimated. The view through the colli-
mation cap should resemble Figure 6e. Make sure all the col-
limation screws are tight (but do not overtighten), to secure 
the mirror tilt.
A simple star test will tell you whether the optics are accu-
rately collimated.

star-testing the telescope

When it is dark, point the telescope at a bright star and accu-
rately center it in the eyepiece’s field of view. Slowly de-focus 
the image with the focusing knob. If the telescope is correctly 
collimated,  the  expanding  disk  should  be  a  perfect  circle 
(Figure 10). If the image is unsymmetrical, the scope is out of 
collimation. The dark shadow cast by the secondary mirror 
should appear in the very center of the out-of-focus circle, like 
the hole in a donut. If the “hole” appears off-center, the tele-
scope is out of collimation.
If you try the star test and the bright star you have selected is 
not accurately centered in the eyepiece, the optics will always 
appear out of collimation, even though they may be perfectly 
aligned. It is critical to keep the star centered, so over time 
you will need to make slight corrections to the telescope’s 
position in order to account for the sky’s apparent motion.

appendix B:  

cleaning the optics

cleaning lenses

Any quality optical lens cleaning tissue and optical lens clean-
ing fluid specifically designed for multi-coated optics can be 
used to clean the exposed lenses of your eyepieces or finder-
scope.  Never  use  regular  glass  cleaner  or  cleaning  fluid 
designed for eyeglasses

10

Figure 8. 

Adjust the tilt of the secondary mirror by loosening or 

tightening the three alignment screws with a small Phillips head 
screwdriver.

Figure 9. 

Loosen one screw on the back of the optical tube one 

full turn and tighten the other screw “in the set” until tight to adjust 
the primary mirror.

Содержание 9883

Страница 1: ...eptional Consumer Optical Products Since 1975 Customer Support 800 676 1343 E mail support telescope com OrionTelescopes com Corporate Offices 831 763 7000 89 Hangar Way Watsonville CA 95076 Orion Spa...

Страница 2: ...r II bracket Micro motion rod attachment post Yoke Azimuth lock knob Tripod leg Accessory tray Leg lock knob Eyepiece Focuser Altitude micro motion control thumbwheel Yoke knob Altitude lock knob Alti...

Страница 3: ...d a flat head screwdriver All screws should be tightened securely to eliminate flexing and wobbling but be careful not to over tighten or the threads may strip Refer to Figure 1 during the assembly pr...

Страница 4: ...Remove the two metal thumbnuts from the optical tube Place the bracket of the EZ Finder II on the tube so that the holes in the bracket slide over the two threaded posts The EZ Finder should be orient...

Страница 5: ...ttery provides the power for the diode To use the EZ Finder II turn the power knob clockwise until you hear a click indicating that power has been turned on With your eye positioned at a comfortable d...

Страница 6: ...s One good way to tell if conditions are good is by how many stars you can see with your naked eye If you cannot see stars of magnitude 3 5 or dimmer then conditions are poor Magnitude is a measure of...

Страница 7: ...with your telescope You should be able to see bands on Jupiter the rings of Saturn craters on the Moon the waxing and waning of Venus and many bright deep sky objects Do not expect to see color as yo...

Страница 8: ...able to ferret out more and more subtle details and structure To find deep sky objects in the sky it is best to consult a star chart or Planisphere These guides will help you locate the brightest and...

Страница 9: ...h the telescope pointed toward a bright surface such as white paper or wall Placing a piece of white paper in the tele scope tube opposite the focuser i e on the other side of the secondary mirror wil...

Страница 10: ...this process on the other two sets of collimation screws if necessary It will take a little trial and error to get a feel for how to tilt the mirror in this way When you have the dot centered as much...

Страница 11: ...To clean the primary mirror carefully remove the mirror cell from the telescope To do this you must loosen the three screws from the end of the optical tube that are flush with the end of the tube Co...

Страница 12: ...tsonville CA 95076 If the product is not registered proof of purchase such as a copy of the original invoice is required This warranty does not apply if in Orion s judgment the instrument has been abu...