LifePort Kidney Transporter Operator’s Manual 755-00002 Rev P
29
Error Message
Probable Cause
Actions
Kidney Not
Connected
Tubeframe Not
Positioned Properly
Leaking Cannula or Artery
Kidney Not Connected
1. Check Tubeframe and Locking Arm position.
2. Under aseptic conditions, visually inspect kidney
and cannula and correct all leaks under aseptic
conditions.
If problem is not resolved, call Organ Recovery Systems
24/7 Perfusion Helpline.
Load Perfusion
Circuit
The Tubeframe Not Properly
Installed or Latched
1. Check Tubeframe and Locking Arm position.
2. Check Pressure Sensor Cable connection.
If problem is not resolved, call Organ Recovery Systems
24/7 Perfusion Helpline.
Low Battery
Less Than 4 Hours Battery
Life Remaining: 2 Hours
Perfusion Plus Additional 2
Hours Monitoring
1. Plug into external power supply.
2. Hot swap batteries for charged batteries.
If problem is not resolved, call Organ Recovery Systems
24/7 Perfusion Helpline.
Motor Current
Failure
LifePort Not Responding
Normally
Call the Organ Recovery Systems 24/7
Perfusion Helpline.
Near Freezing
Incorrect Coolant
Environmental
Conditions Too Cold (Ice
Container temperature has
dropped below 0.1°C)
1.
Check that ice and water only are used to fill
Ice
Container.
2. Move LifePort into a warmer environment.
If problem is not resolved, call Organ Recovery Systems
24/7 Perfusion Helpline.
Occlusion
Unexpected Pressures During
Infuse Mode
1. Under aseptic conditions, check that Infuse Line is
unobstructed.
2. Under aseptic conditions, ensure there are no
blockages or twists in the artery.
If problem is not resolved, call Organ Recovery Systems
24/7 Perfusion Helpline.
POST Failure
Internal Error
1. Remove all power to LifePort Kidney Transporter:
remove all four (4) LifePort batteries and unplug from
AC Mains.
2. Restore power to LifePort and press Power button.
If problem is not resolved, call Organ Recovery Systems
24/7 Perfusion Helpline.
Pressure Sensor
Failure
Pressure Sensor
Disconnected
Reconnect the Pressure Sensor.
If problem is not resolved, call Organ Recovery Systems
24/7 Perfusion Helpline.
Pressure Sensor
Endpoint Error
LifePort Kidney Transporter
Unable to Set Pressure Alert
Levels
1. Press
STOP
to clear alert.
2. Press
INFUSE
to re-enter Infuse Mode.
If problem is not resolved, call Organ Recovery Systems
24/7 Perfusion Helpline.
Purge Bubbles
Automatic Wash Cycle during
Infuse Mode
Possible Air Leak
1. Run LifePort Kidney Transporter in Wash Mode.
2. If this is a persistent error, check Perfusion Circuit for
cracks, leaks, and/or loose fittings.
If problem is not resolved, call Organ Recovery Systems
24/7 Perfusion Helpline.