LifePort Kidney Transporter Operator’s Manual 755-00002 Rev K
54
Power On Self Test (POST)
On Each power up, LifePort Kidney Transporter performs a Power On Self Test or “POST”. The CPU
within LifePort Kidney Transporter checks its memory functions, temperature sensors, air Bubble
Detectors and its internal failure routines. In the unlikely event that one of these tests fail, LifePort
Kidney Transporter will display “POST failure” and it will list the POST error message as shown in the
table below. Should one of these errors occur, remove all power to LifePort Kidney Transporter by
reinstalling the batteries and the mains Power Cord. If the POST message continues to display, call the
Organ Recovery Systems Perfusion Helpline.
POST Error Message
Indication
POST AT error
Ice Bucket temperature error
POST failure
Power on self test failed
POST Flash error
Integrity check of Flash memory
POST HPS error
High pressure shutdown error
POST IB error
Infusion Bubble Detector error
POST MD error
Motor drive error
POST MOC error
Motor overcurrent shutdown error
POST PS error
Pressure Sensor error
POST PT error
Perfusion temperature sensor error
POST RAM error
Integrity check of Read Only Memory
POST ROM error
Integrity check of Read Only Memory
POST UB error
Upstream Bubble Detector error
POST WD error
Watchdog error