Page 1 of 2
INSTRUCTIONS FOR USE
RAYCAST®HIGH PRECISION
BASE PLATE-TO-COUCH TOP
FIXATION DEVICES
2-PIN BAR FOR INTERLOC
COUCH 50.0 CM
(MRI COMPATIBLE)
Article No. :
32207
A.
GENERAL PRODUCT INFORMATION
This product is a fixation device to attach Raycast
®
High
Precision base plates to Interloc couch tops with a width of 500
mm.
B.
PRODUCT DESCRIPTION
This 2-pin bar has two positioning elements that can be inserted
in the half-moon cut-outs on both sides of the Interloc couch. It
can be used to index hardware (base plates, shoulder
positioning device, leg indexing system, etc.) directly or via an
adapter plate to the couch. For information on the compatible
products, please refer to the respective ‘instructions for use’.
C.
POSITIONING ON A COUCH
The two round positioning elements of this fixation device fit
directly in the cut-outs of an Interloc couch.
To position this 2-pin bar on such a couch, gently push the bar
with its two indexing elements into a set of couch cut-outs. The
bar itself should now be flat on the surface of the couch. Always
make sure that the bar is well positioned on the couch before
indexing the other hardware to the 2-pin bar.
Orfit hardware is mounted on the 2-pin bar by simply pushing it
over the 2 middle pins.
When an indexing pin needs to be replaced simply use a Phillips
head screwdriver to do so.
D.
STORAGE
Always store the product in a safe place to prevent it from
getting damaged.
Store the product b10°C (50°F) and 40°C (122°F).
E.
PROPERTIES
E.1 Physical Properties
Dimensions : L 519 mm x W 25.5 mm x H 29 mm
Weight: 0.1 kg
E.2 Material Properties
Material POM (pins/blocks):
1,4 gr/cm³
Material PC (screws):
1,2 gr/cm³
Material Glass Fibre (bar):
1,8 gr/cm³
This product is made entirely of durable plastics and is MR-safe.
Warning: keep metal wrenches which might be supplied in the
package out of the MRI room.
E.3 Dosimetric Properties
Attenuation factor (± 0.15 %)
6 MV
2.67 %
15 MV
2.00 %
H
2
O equivalence
7.72 mm
Note: Use these numbers as a guidance only. Perform the
measurements again in your department to verify these results.