![OREX AA095042-00 Скачать руководство пользователя страница 69](http://html1.mh-extra.com/html/orex/aa095042-00/aa095042-00_service-manual_1650368069.webp)
5.15
Replacing the Plate Size Sensor
1. Remove the cover. Refer to Section 5.1.1, page 37.
2. Locate the Sensor and open its two Allen screws.
3. Cut the tie-wraps that route the sensor cable, and follow the cable
up to the Sensor Board. Disconnect the cable from the Sensor
Board.
4. Connect the new Sensor to the Sensor Board, and route the
Sensor cable to the appropriate location on the system, using tie-
wraps.
Service Operation
Page
64 of 64
Содержание AA095042-00
Страница 9: ...3 2 System Block Diagram Component Description Page 4 of 4...
Страница 10: ...3 3 USB Board 3 3 1 Location 3 3 2 Component Component Description Page 5 of 5...
Страница 12: ...3 4 Motion Board 3 4 1 Location 3 4 2 Component Component Description Page 7 of 7...
Страница 15: ...3 5 Sensor Board 3 5 1 Location 1 3 5 2 Component 2 Component Description Page 10 of 10...
Страница 18: ...3 6 2 Component Destination J300 on USB card 3 7 Laser Board 3 7 1 Location Component Description Page 13 of 13...
Страница 23: ...3 10 Step Motor Slide 3 10 1 Location 3 10 2 Component Component Description Page 18 of 18...
Страница 25: ...3 11 Step Motor Carriage 3 11 1 Location 3 11 2 Component Component Description Page 20 of 20...
Страница 27: ...3 12 Power Supply Assembly 3 12 1 Location 3 12 2 Component Component Description Page 22 of 22...
Страница 29: ...3 13 Erase Lamps Sensor 3 13 1 Location 3 13 2 Component Component Description Page 24 of 24...
Страница 35: ...3 18 Erase Lamps 3 18 1 Location 3 18 2 Component Component Description Page 30 of 30...
Страница 37: ...3 19 Inverters Assembly 3 19 1 Location 3 19 2 Component Component Description Page 32 of 32...
Страница 43: ...5 1 2 Service Panel 1 Open the back panel Philips screws Service Operation Page 38 of 38...
Страница 60: ...4 Disconnect connectors J213 J211 and U211 from the Motion Board Service Operation Page 55 of 55...