State of the memory media
Overwiev of the memory media
When the storage is full at 80% or higher, the icon colour changes to yellow, when the storage
exceeds 90% of its capacity, the icon colour changes to red. If the recording is in progress and
there is no error condition, the icon colour is green. If the recording is not turned on, the icon
colour is white. The logged-in user will click on the icon to access the media overview.
State of the recorder
The recorder always operates in one of its four states (further details in the chapter Errors and warnings).
„
OK
“ - device has not detected any function problem
„
Warning
“ - device outside the specified values, but without consequences on its functioning.
This state can be caused for instance by an undervoltage greater than 10%, by filling the
memory in excess of 80%, and by many other causes. If the device returns within the specified
values, „Warning“ state will disappear and the log record will be carried out.
„
Error
“ - device outside the specified values, but with possible consequences on its functioning.
This state can be caused for instance by an undervoltage greater than 20%, by filling the memory
in excess of 90%, and by many other causes. At the „Error“ state an error window is displayed,
through which you can view the errors. If the device returns within the specified values, „Error“
state will disappear and the logrecord will be carried out.
„
Critical error
“ - device outside the specified values with consequences on its functioning. This
state can be caused for instance by an undervoltage greater than 50% or by filling the memory
up to 100 % so that it is not possible to make records. At the „Critical error“ state an error window
is displayed, through which you can view the errors. If the device returns within the specified
values, both „Critical error“ and the error window remain displayed until a confirmation
(acknowledgment) of the „Critical error“ is done. After confirmation the log record will be carried
out.
Date and time of each error is recorded in the log
!