In this Manu step the 100mA Output Calibration Current can be activated.
The 100mA Output Current
is an Option and has to be ordered from Orbit Controls AG, Switzerland.
Further menu steps contain the correcting factor of the battery charge, temperature correction of the
junction for Thermocouples, storing of new calibration values and recall of the factory settings. The last
position will be used when during the attempt of calibration errors have been done or the calibration
incorrect terminated.
The menu steps contain
- Correction of the junction temperature
- Battery charge display correction
- Restore of factory calibration values
- Backup of new calibration values.
Standard instruments are 0 … 20mA. The output option 100.00mA
has to be ordered at the factory.
The temperature of the output terminals will be measured with
external thermometer. The correction will be done in
Set to
.
Terminate and store with OK.
Check the voltage with external V-meter when the battery is fully
charged and enter the value. Terminate and store with OK.
Restore:
Recall of factory settings.
Backup:
Store new calibration values.