Chapter 4 - Capture images
4.9.2 Artifacts - Eye movements
The patient’s eye may move during the scan. Eye movements can result in a shift or shearing in the image,
and may be visualized by a break in the blood vessel patterns or optic disc. The image should be
recaptured if the eye movement is significant, or if it affects the scan of the fovea or optics disc. Very small
eye movements can be ignored if they do not affect important structures such as the fovea or optic disc.
The Optos software automatically detects possible eye movements (see
page 46). You should replay the frames and assess if the movement is significant.
FIGURE 17: Possible movement detected.
4.9.3 Artifacts - Blinks
The patient may blink move during the scan. Blinks can result in blank B-Scans. The image should be
recaptured if the blink occurs near the fovea or optic disc. Blinks can be ignored if they does not affect
important structures such as the fovea or optic disc.
The Optos software automatically detects possible blinks (see
on page 46). You
should replay the frames and assess if this is significant.
FIGURE 18: Possible blink detected.
Part Number: G110230-003-ENG
Page 49 of 84
Copyright 2019-2021, Optos plc. All rights reserved.
English
Содержание P200TxE
Страница 1: ...User Guide Optos OCT Ophthalmoscope P200TxE G110230 003 ENG...
Страница 4: ...Page 4 of 84 Part Number G110230 003 ENG English Copyright 2019 2021 Optos plc All rights reserved...
Страница 52: ...Page 52 of 84 Part Number G110230 003 ENG English Copyright 2019 2021 Optos plc All rights reserved...
Страница 58: ...Page 58 of 84 Part Number G110230 003 ENG English Copyright 2019 2021 Optos plc All rights reserved...
Страница 64: ...Page 64 of 84 Part Number G110230 003 ENG English Copyright 2019 2021 Optos plc All rights reserved...
Страница 78: ...Page 78 of 84 Part Number G110230 003 ENG English Copyright 2019 2021 Optos plc All rights reserved...
Страница 84: ...G110230 003 ENG English...