different phenomena such as attenuation of the fiber, dirty connector mating faces, 

connector misalignment, loss of splices and other issues such as sharp fiber bends. 

The "end to end" loss test is the most commonly used acceptance test for power loss 

in fiber optic links. The test is based on measurement the power difference at the 

input and output of the link. 

The PM 215 and LS-800/LS-215 are used for this test, where the light source (LS-

800/LS-215) acts  as  a  transmitter and the optical power meter (PM 215) as a 

receiver. 

 

The "end to end" test includes two steps:  

1.

 

Setting the reference 

2.

 

Measurement the loss 

 

10.2

 

Method 6 

 

For method 6 two master cords are used to set the reference. Method 6 cancels the 

effects of the master cords and one adaptor for all subsequent measurements. 

 

10.2.1

 

Setting the reference 

 

1.

 

Connect the first master cord to the power meter (PM-215). 

2.

 

Connect the second master cord to the light source (LS-800/LS-215). 

3.

 

Use the master adaptor to connect the two fiber ends (figure 15). 

4.

 

Power on the light source and by pushing [

λ

].select the appropriate wavelength. 

5.

 

Power on the power meter and by pushing [

λ

select the appropriate wavelength 

range. 

6.

 

On the power meter activate the relative power measurement mode - by pushing 

[REL]

7.

 

Set and store the new reference for the selected wavelength - push [REF], 

[CONFIRM]. 

 

 

 

 

Figure 14: Setting reference by method 6 

 
 

 

 

Содержание PM-215

Страница 1: ...entioned herein may be the trademarks of their respective owners OPTOKON a s Cerveny Kriz 250 586 01 Jihlava Czech Republic tel 420 564 040 111 fax 420 564 040 134 WWW OPTOKON COM INFO OPTOKON COM 22 11 2017 PM 215 Pocket Optical Power Meter INSTRUCTION MANUAL Revision 1 0 ...

Страница 2: ...of the respective owners The publisher and the author make no claim to these trademarks While every precaution has been taken in the preparation of this document the publisher and the author assume no responsibility for errors or omissions or for damages resulting from the use of information contained in this document or from the use of programs and source code that may accompany it In no event sh...

Страница 3: ...ys 9 8 1 1 OFF 9 8 1 2 MEM 10 8 1 3 Menu 1 Absolute power measurement mode REL and λ 10 8 1 4 Menu 2 Relative power measurement mode 10 9 Working with the internal memory 11 9 1 SAVE RESULT 12 9 2 VIEW RESULTS 12 9 3 UPLOAD RESULTS 13 9 4 ERASE RESULTS 15 10 Insertion Loss Measurement 15 10 1 Basic theory 15 10 2 Method 6 16 10 2 1 Setting the reference 16 10 2 2 Measurement Loss 17 10 3 Method 7 ...

Страница 4: ...l loss measurement on singlemode and multimode fiber cable at six working wavelengths 2 Features Small size light weight Six working wavelengths Absolute and Relative optical power measurement Internal two level memory structure Capacity up to 100 measurements SmartProtocol SW generation of test reports Data exporter USB port for Charging the battery Data upload to PC Firmware upgrade Built in Li ...

Страница 5: ...tional Master patchcords Master adaptors Universal SFF 1 25 mm adaptor TE ADP 125 Soft carrying case Changeable input adapters TE AGP FC FC adaptor TE AGP SC SC adaptor TE AGP ST ST adaptor TE AGLP FC FC adaptor TE AGLP SC SC adaptor TE AGLP ST ST adaptor TE AGLP LC LC adaptor Compatible with adapter type G Compatible with adapter type GL TE HC 01 Power charging adaptor Soft case TE EVA 215 130x8x...

Страница 6: ...ifying devices with a built in infra red filter to ensure safety During operation testing or maintenance of a fiber optic system never look into an active fiber optic cable Infrared radiation may be present and permanent eye damage can result Avoid direct exposure to the beam Do not activate the laser when there is no fiber attached to the optical output connector Under no circumstances should you...

Страница 7: ...agents Acetone or other active solvents may damage the case The instrument is resistant to normal dust and moisture however it is not waterproof If moisture does get into instrument dry it out carefully before using it again 7 3 Optical connector care Cleanliness will affect the performance of an optical fiber system All connectors and fiber end faces need to be clean prior to testing Clean all co...

Страница 8: ... 4 The unit always starts up with displaying the settings when it was switched off For example if absolute mode measurement wavelength 1550 nm and Auto Off is selected and the unit is switched off after pressing the ON button the unit will start up with the same settings Figure 4 View of display after turn on PM 215 Power meter port with 2 5 mm universal adaptor Operating keys Auto OFF indicator L...

Страница 9: ...y keys for 3s PM 215 is switched off Figure 6 Auto off function and rottate function AUTO OFF Activation and deactivation function Auto Off Press button OFF displayed figure 7 you can using the button enable disable Auto Off The unit will display this symbol function is activated figure 3 If Auto Off is enabled the unit will switch off after 10 minutes of no activity Rottate Rottate function switc...

Страница 10: ...will appear if the unit was switched off in the absolute power measurement mode λ Selects 850 1300 1310 1490 1550 1625 nm wavelengths REL Activates the relative power measurement mode unit dB Recalls the last set reference for a selected wavelength Goes to menu 2 8 1 4 Menu 2 Relative power measurement mode If the relative power measurement mode is on the value of the optical insertion loss in dB ...

Страница 11: ...rnal memory The PM 215 memory has a structured two level organization The results are stored in memory positions MEM in folders called Cable CAB See the table A maximum of 100 results can be stored in the memory although the maximum number of CAB and MEM is 100 This means if 50 CAB folders are used each CAB folder can contain a maximum of two memory positions MEM and vice versa if only two CAB fol...

Страница 12: ...e appropriate CAB and press OK 4 Use or keys for selecting the appropriate MEM and press OK 9 2 VIEW RESULTS 5 Press MEM 6 Use or keys for selecting View and press OK 7 Use or keys for selecting the appropriate CAB and press OK 8 Use or for selecting the appropriate MEM 9 Measure value will be displayed 10 Press OK to return to Menu 1 or 2 ...

Страница 13: ...a PC using the USB cable provided and turn the PM 215 on The PC will automatic install drivers or prompt you to install the drivers for new hardware Software can download from this page http www optokon cz testers manuals These drivers will create a virtual serial com port b Start putty as administrator from installed folder Figure 10 First start of putty c Choose the connection type on Serial Fig...

Страница 14: ... device check control panel device manager Ports LTP and Serial ports Figure 12 Settings of connection e Use or keys for selecting UPLOAD DATA and press MEM Figure 13 Uploaded result from PM 215 The data from this window can be easily copied to any other application Similar process can use for another network clients Memory position Wavelength Stored values Reference Cable folder ...

Страница 15: ...e programs with manuals can download at this page http www optokon cz testers manuals 9 4 ERASE RESULTS This function erase all stored records from your memory Please Use this function only if you will be sure that you don t need recorded data for future using 1 Press MEM 2 Use or keys for selecting ERASE and press OK 3 Confirm function or not 4 After confirmation all memory contents will be erase...

Страница 16: ...ting the reference 2 Measurement the loss 10 2 Method 6 For method 6 two master cords are used to set the reference Method 6 cancels the effects of the master cords and one adaptor for all subsequent measurements 10 2 1 Setting the reference 1 Connect the first master cord to the power meter PM 215 2 Connect the second master cord to the light source LS 800 LS 215 3 Use the master adaptor to conne...

Страница 17: ...hod 7 one master cord is used to set the reference The master cord will be cancelled for all subsequent measurements 10 3 1 Setting the reference 1 Connect the master cord to the power meter PM 215 2 Connect the second end of the master cord to the light source LS 800 LS 215 3 Power on the light source and select the appropriate wavelength by pushing λ 4 Power on the power meter and select the app...

Страница 18: ...ord from the light source 2 Disconnect the master cord from the power meter port 3 Connect the trace to be measured between the power meter port and the master cord that is attached to the light source An extra master adaptor is required 4 The power meter display will report the loss in dB Figure 17 Measurement Loss by method 7 ...

Страница 19: ...PM 215 GL INSTRUCTION MANUAL 19 Note 11 Power loss and decibels ...

Страница 20: ... 6 0 25 0 9 18 7 5 0 32 1 20 6 4 0 40 2 36 9 3 0 50 3 49 9 2 0 63 4 60 2 1 0 79 5 68 4 0 1 00 6 74 9 1 1 26 7 80 0 3 2 00 8 84 2 5 3 16 9 87 7 7 5 01 10 90 0 10 10 00 12 93 7 12 15 84 15 96 8 15 31 62 20 99 0 17 50 12 30 99 9 20 100 00 12 Calibration service center OPTOKON a s Červený Kříž 250 586 01 Jihlava Czech Republic tel 420 564 040 111 fax 420 564 040 134 OPTOKON OPTOKON COM WWW OPTOKON COM...

Отзывы: