MEX>DOF!_X.XXX
MEX>DOF_X.XXX
Command used to collimate beam or get
required divergence value:
+ to increase beam divergence
- to decrease beam divergence
(For. Ex. You can observe that laser beam
beam is diverging too much and it must be
collimated. To do that MEX>DOF_-0.5 or
different values must be sent while observing
the output)
MEX>DOF!_-0.5
MEX>DOF_-0.5
Settings
|
Baud rate - BAUD, Wavelength - WL, Coe
ffi
cients - CMAG,
MEX>BAUD?
MEX>BAUD_baud rate
Command used to get baud rate.
(57600 by default)
MEX>BAUD?
MEX>BAUD_115200
MEX>BAUD!_baud rate
MEX>BAUD_baud rate
Command used to set baud rate. Available
speeds:
115200, 57600, 38400,
19200, 9600, 4800.
Other values are ignored.
MEX>BAUD!_57600
MEX>BAUD_57600
MEX>CWL?
MEX>CWL_XXX.X
Command to get the current working
wavelength in nanometers.
MEX>CWL?
MEX>CWL_532.0
MEX>CWL!_XXX.X
MEX>CWL_XXX.X
Command to set the current working
wavelength in nanometers. The device will
change its current working wavelength only if
it matches one of the design wavelengths!
Design wavelengths can be seen by issuing
an MEX>INFO? Command.
MEX>CWL!_1064
MEX>CWL_1064.0
MEX>CWL!_999
MEX>CWL_1064.0
MEX>CWL!_532.1
MEX>CWL_532.0
MEX>CMAG?
MEX>CMAG_X.XXXXXX_
Y.YYYYYY_Z.ZZZZZZ_U.
UUUUUU_V.VVVVVV_W.
WWWWWW_
x.xxxxxx_y.yyyyyy_z.zzzz
zz_u.uuuuuu_v.vvvvvv_w.
wwwwww
Command used to get the current polynomial
coefficients of curves A (uppercase) and B
(lowercase). These coefficients are sent in
scientific notation with 6 digits of precision
starting with the 0th order and ending with the
5th order coefficient.
MEX>CMAG?
MEX>CMAG_-1.1154e3_8
.
0183e1_-2.0505e0_2.362
0e-2_-1.0279e-4_0.0000_
-1.1154e3_8.0183e1_-2.0
505e0_2.3620e-2_-1.0279
e-4_0.0000
MEX>CMAG!
_X.XXXXXX_Y.YYYYYY_
Z.ZZZZZZ_U.UUUUUU_
V.VVVVVV_W.WWWWW
W_x.xxxxxx_y.yyyyyy_z.z
zzzzz_u.uuuuuu_v.vvvvv
v_w.wwwwww
MEX>CMAG_X.XXXXXX_
Y.YYYYYY_Z.ZZZZZZ_U.
UUUUUU_V.VVVVVV_W.
WWWWWW_x.xxxxxx_y.
yyyyyy_z.zzzzzz_u.uuuuu
u_v.vvvvvv_w.wwwwww
Command used to set the new polynomial
coefficients of curve A (uppercase) and B
(lowercase). These coefficients are sent in
scientific notation with 6 digits of precision
starting with the 0th order and ending with the
5th order coefficient.
MEX>CMAG!
_-1.1154e3_8.0183e1_-2
.
0505e0_2.362e-2_-1.027
9e-4_0_-1.1154e3_8.018
3e1_-2.0505e0_2.362e-2
_-1.0279e-4_0
MEX>CMAG_-1.1154e3_8
.
0183e1_-2.0505e0_2.362
0e-2_-1.0279e-4_0.0000_
-1.1154e3_8.0183e1_-2.0
505e0_2.3620e-2_-1.0279
e-4_0.0000
Information
|
Device state - STATUS, information - INFO, Serial number - ID
Command used to get the current state of the
device. First flag signifies if element actuation
is enables or disabled. Second flag signifies if
automatic target coordinate calculation is
active. Third flag signifies if calculation mode
is inverted or direct. The error byte shows if
any error state is active. Bit number meaning if
MEX>STATUS?
ENA_CON_DIRECT_ERR_
0
MEX>STATUS?
DIS_CON_INVERT_ERR_3
2
Command
Response
Comments
Example usage
User command
Device response
16
Содержание MEX13-HP
Страница 1: ...MANUAL v8 HIGH POWER MOTORISED BEAM EXPANDERS 1x 3x 1x 5x MEX13 HP MEX15 HP KEY for DISCOVERIES ...
Страница 18: ...9 3 Serial communication example in Python 18 ...
Страница 21: ...OPTOGAMA UAB Mokslininku str 2A LT 08412 Vilnius Lithuania 37052194884 sales optogama com www optogama com ...