Vibration Strength(without packing)
There shall be no sign of malfunction after the following vibration test.
Vibration test:
Increase the frequency of the vibration from 10Hz to 100Hz at an accelerated velocity of
19.6m/s2 (2.0 G) for 30 minutes (60 minutes for a cycle) in the non-operating state. Repeat this in each X, Y
and Z direction.
9.6.
Vibration Strength (in individual packing)
There shall be no sign of malfunction after the following vibration test.
Vibration test:
Increase the frequency of the vibration from 10Hz to 100Hz at an accelerated velocity of 19.6
m/s2 (2.0 G) for 30 minutes (60 minutes for a cycle) in individually packaged state. Repeat this in each X, Y
and Z direction.
9.7.
Drop Impact Strength
There shall be no sign of malfunction after the following drop test.
Drop test:
Drop the scanner three times (18 times in total), at each 6 face, from a height of 150cm onto a
concrete floor as shown below.
150cm
Figure 9:
Drop test
9.8.
Drop Impact Strength (in individual packaging)
There shall be no sign of malfunction after the following drop test.
Drop test:
Drop an individually packaged scanner 10 times in total, at any of 1 corner, 3 edges, and 6 faces,
from a height of 150 cm onto a concrete floor.
9.9.
Electrical Specifications
Air discharge
±8 kV max. (No malfunction)
±15 kV max. (No destruction)
Measurement environment
An electrostatic testing device compliant with IEC 61000-4-2 is
used.
Discharge resistance
330
Ω
Charging capacitor
150 pF