5
3.
Installation and Operation
3.1. Installation
3.1.1
Put the body upright on the table steadily.
3.1.2
Put the head into body.
3.1.3
Put out the dust cover from eyepiece tube,and insert available eyepiece into
eyepiece tube.
3.1.4
Turn around the coarse focusing knob to araise the stage.Remove the dust covers
of nosepiece and put the objectives into the holes in nosepiece as per the power of
objectives.
3.1.5
Put the available stage plate into the hole of stage, and insert the clamps into the
small holes of top face of stage.
3.2. Operation
Check the outside power voltage and make sure it is available then put the socket
into pocket.
3.2.1
Turn around the potentiometer then adjust the light to an applicable position.
3.2.2
Select the suitable sample and place it on the stage plate,make sure the observed
face down,and fix the sample
with clips. Turn around the
stage knob to make the sample
into optical path.
3.2.3
Hold the rim of nosepiece
and turn around to put a low
power objective(usually 10×)
into optical path then observe
the specimen with left eyepiece
by left eye and turn coarse
focusing knob to find the image,
then turn slowly fine focusing
knob to make the image clear. Fig.3
After that,put high power objective into the optical path.
3.2.4. Pupil Distance Adjustment and Diopter Adjustment
3.2.4.1 Interpupillary Distance Adjustment (As shown in Fig.3)
Interpupillary distance is different for everyone, so it is necessary to adjust
interpupillary distance before using binocular microscope. Please take the two
eyepiece tubes and turn until the rings observed by two eyes are in superposition
entirely.
3.2.4.2 Diopter Adjustment (As shown in Fig.3)
As focusing for binocular, user should observe right eyepiece by right eye, and make
the image clear by focusing adjustment, then observe the left eyepiece. At the same
time, adjust the diopter ring of the left tube to make the image in left eyepiece clear as