3.

 

Installation and Operation 

3.1. Installation 

3.1.1

 Put the body upright on the table steadily. 

3.1.2

 Put the head into body. 

3.1.3

  Put  out  the  dust  cover  from  eyepiece  tube,and  insert  available  eyepiece  into 

eyepiece tube. 

3.1.4

 Turn around the coarse focusing knob to araise the stage.Remove the dust covers 

of  nosepiece and  put  the  objectives  into  the  holes  in  nosepiece  as  per  the  power  of 
objectives. 

3.1.5 

Put the available stage plate into the hole of stage, and insert the clamps into the 

small holes of top face of stage. 

3.2. Operation 

Check the outside power voltage and make sure it is available then put the socket 
into pocket. 

3.2.1

 Turn around the potentiometer then adjust the light to an applicable position. 

3.2.2

 Select the suitable sample and place it on the stage plate,make sure the observed 

face  down,and  fix  the  sample 
with  clips.  Turn  around  the 
stage knob to make the sample 
into optical path.   

3.2.3 

Hold the rim of nosepiece 

and  turn  around  to  put  a  low 
power  objective(usually  10×) 
into  optical  path  then  observe 
the specimen with left eyepiece 
by  left  eye  and  turn  coarse 
focusing knob to find the image, 
then turn slowly fine    focusing 
knob to make the image clear.                                              Fig.3 
After that,put high power objective into the optical path.     

 

3.2.4. Pupil Distance Adjustment and Diopter Adjustment 

3.2.4.1 Interpupillary Distance Adjustment (As shown in Fig.3)   
Interpupillary  distance  is  different  for  everyone,  so  it  is  necessary  to  adjust 
interpupillary  distance  before  using  binocular  microscope.  Please  take  the  two 
eyepiece  tubes  and  turn  until  the  rings  observed  by  two  eyes  are  in  superposition 
entirely. 
3.2.4.2 Diopter Adjustment (As shown in Fig.3) 
As focusing for binocular, user should observe right eyepiece by right eye, and make 
the image clear by focusing adjustment, then observe the left eyepiece. At the same 
time, adjust the diopter ring of the left tube to make the image in left eyepiece clear as 

Содержание MDJ A13.2602

Страница 1: ... A13 2602 Student Biological Instruction Manual To ensure the safety and obtain satisfactory performance please study this instruction manual thoroughly bef ore start to use the instrument ...

Страница 2: ...CONTENTS 1 Applications 3 2 Specification 3 3 Installation and Operation 5 4 Maintenance 7 5 Troubleshooting 9 ...

Страница 3: ...2 Fig 1 MDJ200 Metallurgical Microscope Fig 2 MDJ100 Metallurgical Microscope ...

Страница 4: ...0X 80X 5X 25X 50X 100X 250X 400X 10X 40X 100X 200X 500X 800X 12 5X 50X 125X 250X 600X 1000X 16X 64X 160X 320X 800X 1280X 2 2 Objectives Obj Mark Obj Type Magnification Numerical Aperture N A Working Distance mm System Colour Ring LPL5X 0 13 0 LPL 5 0 13 16 07 Dry Red LPL10X 0 25 0 LPL 10 0 25 18 48 Dry Yellow LPL20 0 40 0 LPL 20 0 40 8 35 Dry Light Green LPL50 0 70 0 LPL 50 0 70 1 95 Dry Light Blu...

Страница 5: ...t Green Blue Grey and White grass can be supplied 2 6 Inllumination 12V 20W halogen bulb adjustable brightness 2 7 Nosepiece Quadruple nosepiece 2 8 Head a Monocular Head b Seidentopf Binocular Head Inclined 45 Interpupillary Adjustable Distance Is 48 75mm Diopter Adjustable Range 5 2 9 Coaxial Coarse Fine Focusing System Coaxial coarse fine focusing knobs coarse stroke 42 4mm rotation actual coar...

Страница 6: ...ound the stage knob to make the sample into optical path 3 2 3 Hold the rim of nosepiece and turn around to put a low power objective usually 10 into optical path then observe the specimen with left eyepiece by left eye and turn coarse focusing knob to find the image then turn slowly fine focusing knob to make the image clear Fig 3 After that put high power objective into the optical path 3 2 4 Pu...

Страница 7: ... image fills 70 80 of the objective aperture After the above points please choose available objective eyepiece and filter as per the sample and adjust aperture iris Fig 5 bright focusing knobs to get the clear image 3 2 7 Usage of the tension adjustment ring Shown in Fig 6 The tension adjustment ring can adjust the tension of the coarse and fine focusing unit to prevent the stage from sliding down...

Страница 8: ...bsorbent cotton or cleaned by absorbent cotton lens paper with the mixture of alcohol and ether proportion 1 4 4 1 2 Alcohol and ether all are burnt early please take them away from fire Be careful for turn on and off power 4 1 3 Don t clean painted metal and galvanizing metal with organic solvent such as alcohol ether or the mixture of the both Silicon cloth or soft cleaning preparation is sugges...

Страница 9: ...over 4 3 5 Pull out the bulb should be replaced hold a new bulb with silk cloth to avoid fingerprint and dust affect bulb brightness and service life and insert fully the contact pins into the bulb socket 4 3 6 Close the lamp housing cover and screw the knurled thumb screw After working for above 10 hours continuously better cut off the microscope about 30 minutes 4 4 Replacement of fuse 4 4 1 Cut...

Страница 10: ... meet the requirement Replacing bulb Brightness isn t adjusted correctly Adjust Rotation potentiometer Objective isn t in correct position Make the objective in correct position The size of iris aperture is too small Adjust the size of iris aperture Lens objective eyepiece Condenser light collector has dust Clean it Image isn t clear contras or definition isn t enough Size of iris aperture isn t p...

Страница 11: ...MDJ Metallurgical Microscope ...

Отзывы: