(1) Sample preparation & installation (sample height adjustment)
1.
Open the door with the vacuum completely released.
2.
Turn the bracket fixed on the control shaft counterclockwise.
3.4. Unscrew the sample stage fixed on the bracket and install the prepared sample.
5. Fix the height of the sample and adjust the height by W/D JIG. (W/D refers to working
distance)
- Rotate the upper and lower parts of the two nuts mounted on the bracket, the height of
the upper edge of the sample and the lower part of the limit arm of the clamp. After
completion, tighten the fixing nuts in opposite directions without changing the height of the
sample.
※
If the sample is relatively high, a nut can be removed to adjust the height.
Fix the stage on the bracket and turn it clockwise to fix it in place.
6. Slowly close the cabin door, press the vacuum button at the same time to vacuum, wait
until the cabin door is sucked and let go.