cold light source(7).   

2.View-speed  adjustment  :  Fix  up  eyepiece,  drive  binocular  head(16)  to  adjust 

interpupillary distance, shown as Fig.2 

 

 

 

Fig.2 

3.Horizontal  working  table:  Place  some  specimen  on  two  table  plates,  separately 

drive left and right lifting knob(5) to keep the image clear. For moving left and right 

the stage, rotate the knob(21); for moving forward and backward the stage , rotate the 

knob(22). 

4.Adjustment  of  separation  line:  The  separation  line  should  be  a  fine  and  black 

straight one. Turn the separation line adjusting knob(10) to move the separation line, 

you  may  get    the  left  and  right  single  viewfield  observation,  cutting-image    and 

superimposed-image  comparison.  When  black  and  white  line  are  wider  ,  rotate  the 

black  screw  at  the  right  of  the  knob(10),  when  width  are  not  identical  ,  rotate  the 

black one at the left of the knob(10). 

  

 

 

Fine and black                  Wider black line                          Width are not identical 

Straight line

 

                           

and white line   

Fig.3 

5.

Magnification Adjustment: For in use, both left and right objective should be at the 

same  magnification  .  To  rotate  the  six-holes  nosepiece  (8)  to  select  the  required 

Содержание A18.1828-LCD

Страница 1: ...A18 1828 LCD LCD Comparison Microscope Instruction Manual To ensure the safety and obtain satisfactory performance please study this instruction manual thoroughly be fore start to use the instrument...

Страница 2: ...tion With our success in solving the technical problem of the adjustability of its separation line width its quality has therefore reached the international advanced standard This model is an ideal in...

Страница 3: ...Horizontal linkage knob 5 Working table lifting knob 6 LED light source 7 Objective 8 Six holes nosepiece 9 Separation line window 10 Separation line adjusting knob 11 Video recording tube 12 CCD opti...

Страница 4: ...bes insert the required eyepieces into the tubes 3 Plug one end of the power line into the socket on the base the other to the socket with 110V 220V ground protection 4 After the above steps are compl...

Страница 5: ...n line The separation line should be a fine and black straight one Turn the separation line adjusting knob 10 to move the separation line you may get the left and right single viewfield observation cu...

Страница 6: ...m through eyepieces until two reticles scale can completely coincide If the left and right scales are not matched turn the magnification correcting knob until the reticles images are sharp and then mo...

Страница 7: ...rvation draw out the lever for photographing Fit the back of camera into the adapter 13 and then insert to the instrument Observe the object through viewframe adjust both left and right stages to obta...

Страница 8: ...enances The instrument must be kept in shade cool and dry place free of dust acid alkali and vapor As soon as work is finished the dustcover should be put on it The instrument must be kept clean The d...

Страница 9: ...ective 0 5x 1x 1 5x 2x 4x 6x 2 pcs each 5W LED Light 2 pcs Auxiliary Lens 0 4X 2 0x 2 pcs each 3 0 M CCD Built in Main Body 1 pcs Polarizer Analyzer 2 pcs each Power cable 1 pcs Coaxial light device 2...

Страница 10: ...www cnoec com cn Page 10 of 10 sale cnoec com...

Отзывы: