-7-
4.2 Installation
4.2.1 Please confirm the installation condition meets 4.1;
4.2.2 Put out the main body and place it in table, and loose the main body
fixing screw, then put the cost cover out;
4.2.3 Put out the intermediate, fix it into the main body, then tighten the
main body fixing screw; If it is Transmitting & Reflecting Polarizer
Microscope, put out the reflecting illumination unit to fix it.
Installation of reflecting illumination unit: Put out the halogen lamp
house, fix it onto the back of reflecting illumination unit by the fixed screw,
connect the power wire of halogen lamp house into the power output
socket of power house.
4.2.4 Fix the head into the intermediate;
4.2.5 Insert the eyepieces into the tubes;
4.2.6 Put out the dust cover of the nosepiece, and turn the coarse focusing
knob to lower the stage, and find the objective hole with yellow mark in
the nosepiece. Fix the 10X objective into the hole, and turn the nosepiece
clockwise, fix the other objectives as per the power.
4.2.7 Loose the condenser lock thumb and fix the condenser.
5. Operation
5.1 Turn the power switch and the potentiometer to adjust light to be
available;
5.2 Fix the sample on the stage, and move it into the path;
5.3 Turn the nosepiece to put 10X objective into light path, and turn the
focus knobs to get clear image.
5.4 Confirm the polarizing vibrancy direction
The polarizing vibrancy direction has set to be at west-east in factory
when the scale of the polarizing is 0°.
5.5 Check polarizing and analyzer
The field should be dark completely (when there is no sample) when the
scale of polarizer and analyzer is 0°. Please check the position of the