11 

 

Fig.13 

Fig.15 

Fig.14 

4. Polarized Light Observation               

 

 

                     

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 
 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 
 
 
 

 

 

 

 

                 

 
   
 
 
 
 
 
 
 
 

       

 

4-1 Adjustments before observation 

4-1-1 Focus 

Use  the  10×objective  to  focus.  To  avoid  the 
objective  touch  with  the  specimen,  you  should 
raise  the  mechanical  stage  and  let  the  specimen 
close  to  the  objective,  then  slowly  separating 
them  to  focus.  The  operator  can  converse  turn 
the  coarse  focus  knob  to  get  the  specimen 
down  ,and  search  images  in  the  10×ocular 
simultaneously, then use the fine knob to focus. 
At  this  moment,  you  can  replace  other 
magnification  objectives  safely,  and  focus 
without the risk of destroying the specimen. 

The  tension  of  the  coarse  focusing  knob  has  been 
adjusted  before  leaving  factory.  If  it  is  too  loose, 
for  example,  the  stage  decline  slightly  due  to 
self-weight,  the  tension  adjustment  collar  can  be 
adjusted. 

4-1-2 A

djusting the optical axis (Fig.13) 

The  revolving  nosepiece  is  centerable.  Before 
leaving  the  factory,  the  optical  axis  of  the 
objective  and  the  center  of  stage  has  been 
adjusted  coaxial,  please  do  not  adjust  them 
unless necessary.

 

★ 

Make sure the 10X objective is inserted into the 
primary hole on the centerable revolving 
nosepiece. 
Perform the following adjustments if the optical 
axial and the center of stage is not coaxial. 

1. Focus on the specimen by 10x objective and look 

for an easily recognizable detail

  in the field. 

Move this detail in the center of the eyepiece cross 
lines. 

2. When the stage is rotated, the detail moves in circle 

. Manipulate the two stage centering knobs

to 

coincide the imaginary center of the circle 
circumscribed by the detail with the intersection 
of the eyepiece cross lines. 

 

 

 

 

 

Содержание A15.1018

Страница 1: ...5 1018 Polarizing Microscope Transmit Light Instruction Manual To ensure the safety and obtain satisfactory performance please study this instruction manual thoroughly be fore start to use the instrument ...

Страница 2: ...1 CONTENTS CONTENTS 1 USER NOTIECE 2 1 NOMENCLATURE 4 2 ASSEMBLY 5 2 1 Assembly Diagram 5 2 2 Detailed Assembly Procedure 6 3 USING THE CONTROLS 9 4 POLARIZED LIGHT OBSERVATION 11 5 SPECIFICATIONS 13 ...

Страница 3: ...y cool designated bulb 24V 100W halogen lamp 7 Use the factory supplied power cord please II Maintenance 1 All the lenses have been well checked and adjusted It is forbidden to disassemble them yourself 2 The nosepiece and coarse fine focus unit have a compact and precise frame so please don t disassemble them as possible as you can 3 Keep the instrument clean wipe dust regularly and be attention ...

Страница 4: ...the microscope That will affect the function or decline the performance of the microscope 8 Place the instrument in a cool dry position After using the microscope remember to cover it with dust helmet Do wait for the lamp house cooling completely before cover ...

Страница 5: ...4 1 NOMENCLATURE Video port Eyepiece Objective Revolving round stage Lamp housing Trinocular viewing head Reflected illumination system Bertrand lens unit Main body ...

Страница 6: ... various components The numbers indicate the order of assembly When assembling the components make sure that all parts are free of dust and dirt and avoid scratching any parts or touching glass surfaces Keeping the hexagon spanner well when change the spare parts you will use it ...

Страница 7: ... and the guide board fixed together 2 2 2Mouting Revolving Round Stage 1 Loosen the centering knobs 2 Position the stage with the venire scale in front and carefully lower the stage onto the round dovetail on the stage bracket with the positioning pin on the stage aligned with the groove on the front of the stage bracket then loosely tighten the centering knobs Turn the stage to check it is mounte...

Страница 8: ...rstly then replace with higher magnification objective if necessary When replacing the objective slowly turning the nosepiece until you hear clicked that means the objective enter the required position the light path center 2 2 4Mounting reflected illumination system Fig 6 1 Mount the Illumination System into the round dovetail on the head of the microscope body properly and use the Allen screw dr...

Страница 9: ...crew down the bolt to fix it For transmitted light illumination the power jack is used For reflected light illumination the power jack is used 2 2 10 Mounting Polarizer and Analyzer 1 Insert the polarizer into its slot with the silkscreen surface toward the front and make it in the light path center 2 push the analyzer into its slot NOTE the polarizer and analyzer are engaged when pushed in to the...

Страница 10: ...g Compensator Vernier scale 平台支紧螺钉 Polarizing 调节旋钮 起偏器转环 勃氏镜转盘 Analyzer Bertrand lens unit Bertrand lens focusing plate Compensator clamping knob Polarizer clamping knob Polarizer rotation ring Polarizing ruler adjustment knob Stage rotation clamping knob ...

Страница 11: ...op position the specimen is moved 45 to its diagonal position To release the 45 click stop function push back the lever 3 2 3 Using the Bertrand Lens By revolving the Bertrand lens dial the Bertrand lens can be selected At the O position the lens is removed from the light path At the B position the lens is engaged 3 2 4 Focusing the Bertrand Lens During conoscopic observation to focus the conoscop...

Страница 12: ...the stage decline slightly due to self weight the tension adjustment collar can be adjusted 4 1 2 Adjusting the optical axis Fig 13 The revolving nosepiece is centerable Before leaving the factory the optical axis of the objective and the center of stage has been adjusted coaxial please do not adjust them unless necessary Make sure the 10X objective is inserted into the primary hole on the centera...

Страница 13: ...ge the polarizer and analyzer for extinction position 2 Swing the condenser top lens into the light path 3 revolve the Bertrand lens dial to B position to engae the lens into the light path 4 open the aperture iris diaphragm to largest size 5 revolve the focusing dial of the Bertrand lens to focus on the conoscopic image 3 Moving the specimen only move a new specimen detail into the cross line cen...

Страница 14: ... eyepiece 10 10 10 10 10 10 objective 5 10 20 40 50 100 Total magnification 50 100 200 400 500 1000 5 Rotatable stage Diameterф170mm 360 scale Minimum reading 6 by means of vernier scale 6 Coaxial Coarse and Fine Focusing System Sensitivity and Graduation of Fine Focus 0 001mm Focusing movement range 32mm 7 Polarizer can be swung out 8 Analyzer 0 360 rotatable 9 Bertrand lens can be removed from o...

Страница 15: ...able quintuple Test plate compensato r Λtest plate first order red 1 4λtest plate Quartz wedge test plate Empty test plate filter ND25 ND6 fitler Blue yellow green frosted illumination 24V100W halogen lamp Spare lamp eyepiece 10 plan eyepiece 2 2 2 10 Crisscross eyepiece 10 eyepiece with scale of crosshair 10 mesh eyepiece Strain free objective Plan 5 Plan 10 Plan 20 Plan 40 Plan 100 oil Plan 50 P...