                                                                        A13.0905     

2-2 Assemblage step 

2-2-1 Set lamp holder 

b                                                                            Make sure the pin 

 and 

 facing to the hole 

 and 

, then lock. (Fig. 2) 

 

LED bulb: 3W 

 Halogen bulb: 6V30W (

Philips 5761) 

 

2-2-2 Set or replace bulb 

Use clean glove or gauze to hold the bulb and 

make  sure  the  bulb  pin 

 facing to hole 

 

and lock. Please make sure the bulb should be 

vertical. (Fig. 3) 

 

 

Do not touch the bulb by finger directly. 

C                                                                                    lean  by gauze  if  there  is  finger  print on 

bu                                                                                  bulb. 

 

Whenever  replace  the  bulb,  please  cut 

off  the  power  first  and  wait  the 

illuminator part cool down. 

 

2-2-3 Set objectives

 

Turn the coarse knob to make the nosepiece to 

lowest.  Through  the  hole  on  stage  to  set  the 

objectives 

  on  nosepiece    from  low 

magnification  to  high  magnification  with  the 

direction. (Fig. 4)

 

 Clean the objectives regularly. It  is very 

sensitive  of  dust  for  objective  of  invert 

microscope. 

  Please  use  low  magnification  objective, 

such  as  5X  or  10X  to  find  image,  then  use 

high one for observation. 

 Make sure to hear click when turning to 

show the objective on correct position.   

 

 

Fig.3 

 

Fig.4 

Fig.2 

 

Содержание A13.0905

Страница 1: ...E200 series To ensure the safety and obtain optimum performance and to familiarize yourself fully with use of this microscope we recommend that you study this manual thoroughly before operating the mi...

Страница 2: ...umination 11 3 2 Focusing 11 3 3 Adjust tension of focus 11 3 4 Adjust diopter 12 3 5 Adjust inter pupillary distance 12 3 6 Use filter 12 3 7 Adjust field diaphragm 13 3 8 Adjust aperture diaphragm 1...

Страница 3: ...he high temperature damp dusty or acute shake place Make sure the worktable is horizontal Following environment is required Indoor temperature 5 40 Max relative humidity 80 1 3 To move the microscope...

Страница 4: ...ganic solution to clean the surfaces of the microscope especially for plastic parts Please use the neutral detergent if necessary 2 3 If the microscope damped by the liquid during working please cut o...

Страница 5: ...ec com 1 Nomenclature A13 0905 Mechanical stage Eyepiece Knob for adjustment tension Coarse knob Fine knob Analyser Knob for brightness Pole for field diaphragm Pole for aperture diaphragm Polarizer s...

Страница 6: ...www optoedu com 6 18 sale cnoec com A13 0905 Objective IE200M stand Power switch Trinocular head Centering screw for field diaphragm Diopter adjustment ring Reflected illuminator Nosepiece Lamp holder...

Страница 7: ...e below numbers shows the assemblage order Before assemblage please make sure all parts are without any dust or dirt Be careful during assemblage and do not scrap or touch the optical lens Plug Lamp h...

Страница 8: ...e is finger print on bu bulb Whenever replace the bulb please cut off the power first and wait the illuminator part cool down 2 2 3 Set objectives Turn the coarse knob to make the nosepiece to lowest...

Страница 9: ...d Fig 5 Turn the metal stage to make the V hole direct to the user 2 2 5 Set clips Take out clips from pole Fig 6 Screw pole into hole on mechanical stage then lock by screw nut Set the clips into pol...

Страница 10: ...ator with its face up to hear click Fig 9 2 2 9 Connect power supply Make sure the main power switch is at O OFF position Insert one side of power plug into microscope socket and the other side connec...

Страница 11: ...objective 5X Please make sure the specimen should be vertical with the objective If not please fix the specimen by plasticene Adjust the diopter of right tube to 0 See 3 4 Adjust diopter Observe throu...

Страница 12: ...r Please make sure the diopter of standard one should be at O position before focusing 3 5 Adjust inter pupillary distance When two eyes observe please turn the two prism house with the direction indi...

Страница 13: ...centering screws of field diaphragm to make the diaphragm image to the center Open the field diaphragm slowly if the field inscribed with its image it indicates the diaphragm centered Fig 18 Please i...

Страница 14: ...ower switch is at 0 Off and cut off the power supply Use screwdriver to take off the fuse from fuse holder and replace a new one Fig 21 The rating for fuse 250V 8A 3 11 Mount and use TV device Loose t...

Страница 15: ...ap Fig 23 Mount the photo device as the direction indicated into tri tube then lock on screw Loose the lock screw on photo adapter and take off photo tube Fig 24 Insert 3 2X photo eyepiece into tube T...

Страница 16: ...quintuple nosepiece Focus Fine and coarse coaxial moving range 33mm fine precision 0 001mm Stage Size 180 155mm moving range 75 40mm 12 metal stage 25metal stage for selection Illumination Reflected...

Страница 17: ...analyzer are not at right position Pull the slides in or pull in the light path selection pole dirt inside of field Dirt on specimen Clean specimen Dirt on lenses Clean lenses image blur Slide cover...

Страница 18: ...ere is not power supply Connect the power supply The wires do not connect right Connect it correctly Faulty setting of bulb Set the bulb correctly Bulb burn out Replace bulb Not enough brightness Faul...