Adjust the left diopter again to make left and right images both clear.     

 
 

4. Rotate objective cover

select appropriate objective multiple. 

5.  When  needed,  can  be  screwed  into  auxiliary  objective  at  the  bottom  of  the  lens 

cover. 

6. If feel the torque too tight when using, rotate the focusing knob arduously; Or the 

torque too light, observing head auto-drop down, please use wrench to adjust the torque of 
inside focusing knob to be appropriate.       

Maintenance and Care   

1. Whether the instrument in use or storage, should avoid dust, humidity, cold, heat or 

PH steam.   

2.  Chemicals shall not be placed near the instrument, even more in 

the microscope cabinet (except desiccant).   

3.  When not in use, the whole instrument should be covered by 

dust cover to keep it clean. If there is dust on the lens surface, don't wipe it by hand, only by 
brush.     

4.  If there is dust on the lens surface, can use clean old linen or cotton cloth dipped 

in a small amount of pure alcohol and ether mixture (1:1), or benzoline and xylene to wipe 
gently. Don't touch the lens surface by finger.   

5.  Due to time, grease on gear, rack, sliding groove surface and sliding cylinder will 

bond with dust into dirt or hardened, lead to rotating the coarse knob difficultly, can use 
xylene to clean, then add a little acid-free animal grease or acid-free Vaseline, can't use 
other grease.       

6.  When microscope and eyepiece are not in use, should be covered with cloth or 

glass to prevent dust.     

 

 

 