Use Notices                                                                                      A10.1018 

I

Safety Note 

1.  Carefully open the box, avoid the accessories, dropping to ground and being damaged.   

2.  Do  keep  the  instrument  out  of  direct  sunlight,  high  temperature  or  humidity,  dusty  and  easy 

shaking environment. Make sure the stage is smooth, horizontal and firm enough. 

3.  When  moving  the  instrument,  please  use  two  hands  to  grip  with  the  two  sides  of  the 

microscope body. 

4.  If the bacterium solution or the water splash to the stage, objective or viewing tube, pull out the 

power cord at once, and wipe up the microscope. Otherwise, the instrument will be damaged. 

5.  When  running,  the  lamp  house  and  nearby  parts  will  be  very  hot.  Please  ensure  there  is 

enough cooling room for them. 

6.  Make sure the instrument is earthed, to avoid lighting strike. 

7.  For safety, be sure the main switch is in 

“O”(off) state before replace the halogen lamp or the 

fuse,  then  cut  off  the  power,  and  do  the  operation  after  the  lamp  bulb  and  the  lamp  house 

completely cool. 

8.  Check the input voltage: be sure the input voltage which signed in the back of the microscope is 

consistent with the power supply voltage, or it will bring a serious damage to the instrument. 

II

Maintenance   

1.  All  the  lenses  have  been  well  checked  and  adjusted.  It  is  forbidden  to  disassemble  them 

yourself. 

2.  The  nosepiece  and  coarse/fine  focus  unit  have  a  compact  and  precise  frame,  please  don

’t 

disassemble them as possible as you can. 

3.  Keep  the  instrument  clean,  wipe  dust  regularly,  and  be  attention  to  avoid  contaminating  the 

optical elements especially.   

4. 

The contaminations on the prism, as finger mark and oil, could be gently wiped with a piece of soft 

cloth or tissue paper, gauze which has been immersed in pure alcohol or xylene. 

 

5.  Don

’t use organic solvent to wipe the non-optical elements, when you need to clean, use the 

soft detergent, please. 

6.  When using, if the microscope is splash by liquid, cut off the power at once, and wipe up the 

moisture. 

7.  Do  not  disassemble  any  parts  of  the  microscope.  That  will  affect  the  function  or  decline  the 

performance of the microscope. 

8.  Place the instrument in a cool, dry position. After using the microscope, remember to cover it 

with dust helmet. Do wait for the lamp house cooling completely before cover. 

Содержание A10.1018

Страница 1: ...A10 1018 Professional Dark Field Microscope Instruction Manual Please read this manual carefully before use microscope...

Страница 2: ...voltage which signed in the back of the microscope is consistent with the power supply voltage or it will bring a serious damage to the instrument II Maintenance 1 All the lenses have been well checke...

Страница 3: ...The numbers in the figure show the installation steps Before installing be sure every components is clean do not touch glass surface Keep well with the supplied hexagon wrench When changing the compon...

Страница 4: ...Microscope Body Lamp House Mechanical Stage Support Device 10 Wide Field Eyepiece Infinite Plan Objective Trinocular Viewing Unit Video Port optional...

Страница 5: ...al Stage Support Device Before installing the device be sure to adjust the coarse focus knob Make the guide board see figure 1 down to the lowest position so you can install the mechanical stage suppo...

Страница 6: ...crew down the bolt to fix See figure 5 2 2 3 Installing and Replacing the Lamp figure 6 Please use the specified halogen Lamp 6V30W 1 Hold to the bulb after you wrap it with gauze or other protection...

Страница 7: ...sepiece from the left or the right side figure 9 then push the nosepiece clockwise then place other objectives by the sequence of low to high magnification figure 10 Installing objective this way will...

Страница 8: ...1 into the eyepiece tube until they are against each other The result is showing in the figure 14 2 2 7Installing the Video Port optional Insert the video port figure 12 into the trinocular unit figur...

Страница 9: ...3 Adjustment A10 1018 Diopter Ring Left Fine Focus Knob Left Coarse Focus Knob Clocking Set Condenser Focus Knob Video Port optional...

Страница 10: ...eo Port optional Interpupillary Distance Indicator Right Fine Focus Knob Tension Adjustment Collar Right Coarse Focus Knob Brightness Adjustment Knob Main Switch Field Diaphragm Swing out Condenser wi...

Страница 11: ...direction the voltage decline and the brightness weaken Using the lamp in a low voltage condition will prolong the use life 4 3 Adjust the Tension Adjustment Collar figure 17 The tightness of the ten...

Страница 12: ...ing the objective touch the specimen 4 5 Adjusting the Interpupillary Distance Figure 19 The Interpupillary distance range 48mm 75mm When observing with two eyes hold on the left and right prism holde...

Страница 13: ...troying the specimen 2 When using the video set Pull out the light path selector lever see figure25 observe with both eyes when the image is sharp you can see the pictures directly on the video screen...

Страница 14: ...Lever Figure 24 4 9 Adjusting Field Diaphragm Figure 24 The control for the field diaphragm is a ring used for adjusting the area of field diaphragm When using turn the ring to reduce the field diaphr...

Страница 15: ...he image outline 8 Adjust two screws with the condenser placed make field diaphragm into the center of optical axis 9 Allow adjustment of iris diaphragm and field diaphragm to highest 10 Slowly raise...

Страница 16: ...amp bulb use the specified halogen Lamp 6V30W 2 The edge of the field of view has shadow or the brightness is asymmetry The nosepiece is not in the located position Adjust it into the located position...

Страница 17: ...r move the eyes away to see other things then back into the objective II Mechanical Part 1 The coarse focus knob is hard to run The tension adjustment collar is too tight Loose properly 2 The image ca...

Отзывы: